98 DF-760E
User Interface
Three ranges can be entered in this screen. The range of the primary Full Scale (FS) must
be less than that of the Expanded Full Scale “A” (FS A) which must be less than that of the
Expanded Full Scale “B” (FS B). The analyzer auto-ranges between the three outputs
depending on the current analyzer reading. Relay contacts can then be assigned to signal a
change in range. If only one expanded range is required, rather than two, then the value of
FS B should be set to equal the FS A value. A window as narrow as 10% of the analyzer’s
decades can be set for the full-scale analog output.
Figure 101: Output SetUp Screen
The
In Calibration Relay
can be enabled to signal that a zero or span calibration is in
process. In addition, the user has the option to freeze the analog output or enable the analog
output to update as the calibration progresses.
7.6.9 Oxygen Graph Setup Screen
Figure 102: Oxygen Graph Setup Menu
The
Graph Setup
is used to adjust the time scale on the front display of the analyzer. A
specific time interval in minutes can be chosen for that display. The information on the
display represents current data and will show a history of concentration from the given
time span to the present. The Y-axis of the main data display is auto-ranging.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...