![Servomex NanoTrace DF-760E Скачать руководство пользователя страница 130](http://html1.mh-extra.com/html/servomex/nanotrace-df-760e/nanotrace-df-760e_instruction-manual_1247053130.webp)
126 DF-760E
Service
Depending upon the nature of the application, Delta F suggests verifying the span
calibration of the analyzer approximately every 12 months using a gas with a known level
of oxygen. Span checks can be performed with gases in the range of 0 to 10 ppm.
However, reliable standard gas mixtures are readily available in the 4-7 ppm range.
The zero calibration is most important for applications requiring accuracy below 10 ppb.
For use in applications that are above 10 ppb, the zero calibration is not recommended for
newly installed instruments.
By far the zero calibration is the most important calibration for the NanoTrace Dual
Moisture/Oxygen Analyzer. From a stable zero calibration baseline, oxygen readings
below 1 ppb can be made accurately. In many measurement cases, the accuracy of the
oxygen reading are determined by the quality of the zero calibration. It is important to
check the zero periodically and make appropriate calibration adjustments.
From an initial start-up, the analyzer may take 7 to 14 days to reach a stable zero. After
achieving a stable zero baseline, the analyzer requires periodic zero checks and possibly
adjustments to ensure accuracy. For applications where the process is continuously
monitored, the zero check frequency guidelines in the table below should be used.
Oxygen Zero Check Frequency Guidelines
Typical Reading
Maximum O
2
Impurity
Zero Check Frequency
1.0 ppb and greater
10 ppb
Once every month
0.2 to 1.0
ppb
5
ppb
Twice per month
5 ppb
Twice per month
0 to 0.2
ppb
1 ppb
Once per week
1 ppb
Once per week
Table 11: Zero Check Frequency
Experience with a particular application determines the optimum frequency of zero
checking.
Accurate oxygen readings can be made even though the zero is not completely stabilized,
such as after a start-up or after exposure to high oxygen concentrations. Simply calculate
the difference in concentration between the analyzer output on zero gas and the sample
gas. This comparison should be made over a short time span to avoid errors introduced by
a stabilizing zero.
NOTE
If the Active Zero Offset feature is in use, and the “Cal Zero”
message appears on the display, it is mandatory to do a zero
calibration.
9.3.2 Storage Conditions
If the analyzer is to be stored for extended periods of time, be sure that the temperature of
storage location does not exceed 50° C (122° F). Storage in direct sunlight can cause
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...