52 DF-760E
User Interface
7.5.2.1.3 Do A Manual Zero
Figure 25: Moisture Manual Zero Screen
The manual zero command enables the user to zero the moisture cell in an interactive
manner.
For this purpose, it is necessary to ensure that moisture free gas is entering the sensor.
This can be accomplished by opening the zero gas valves for the moisture cell. See Figure
24. When switching to a gas that is moisture free, it is important to wait a period necessary
to allow the reading to re-stabilize.
NOTE: If oxygen is included in the background gas calculation, see page 103, the on-
board zero gas purifier is automatically prevented from being used. As a result an
adjustment of the moisture zero, must be done manually and the zero gas must be supplied
from the outside through the sample inlet.
After opening the zero gas valve and selecting
Do a Manual Zero
, a screen will appear
which displays a trace of the recent moisture reading. See Figure 25. Observe the trace
until the reading is stable and then press
Enter
. This action will accept the present value
as the new zero setting for the moisture cell and the
Zero Reference
field will be updated.
After this action, the user will be brought back to the main display.
Pressing ESC during the calibration process will abort the action and return the user to the
main display. Pressing Accept will complete the zero process, update the zero log and
return the user to the main display.
Once complete the gas valves must be returned to the original state with process gas going
directly through the sensor. See Figure 24.
A relay is available on the Moisture Analog Output Setup Screen (see page 66) to signal
that a zero calibration is taking place, and the analog output signal can also be frozen or
allowed to update during the calibration process.
7.5.2.2 Do an AutoZero
The AutoZero command enables the user to zero the moisture cell in an automated manner.
In addition, the AutoZero procedure can be scheduled to run automatically by using the
AutoZero Setup procedure described on page 54.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...