8
ECG Recording Analysis
8.2
Resting ECG Recordings
Page 86
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
8.2.2
ECG Filter
The filter is designed to help reduce muscle artefact. The cut-off frequencies and
filter settings are defined in system settings
(see para. 11.11.3, Display Filter, page
132).
Click the
LP Filter
icon
to toggle the filter between the defined frequencies.
When the filter is enabled the icon is highlighted and indicated in the bottom of the
screen.
When the filter is active, possible morphological signal distortion can occur.
8.2.3
ECG Measurements
Click where you wish to commence measurement and drag the mouse to display
measurements in relation to the start point. The measurement remains on the screen
until the mouse is again clicked and a new measurement can be taken. The
measurements calculated are as follows:
• Amplitude in mV
• Time in seconds
• Slope in mV/s
8.2.4
Zoom
Click the zoom icon
in the top left of the screen to display a zoom view above
the selected section.
8.2.5
Amplitude and Speed
The amplitude and speed of the ECG trace is adjusted with the icon in the top left of
the screen.
The amplitude and speed of a waveform can also be changed using the mouse
wheel:
• Position the cursor in any ECG curve area and rotate the
mouse wheel to increase / decrease the
trace amplitude
.
• Press the
Alt key
and rotate the mouse wheel to increase
/ decrease the
trace speed
.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...