![Schiller CARDIOVIT CS-104 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/schiller/cardiovit-cs-104/cardiovit-cs-104_user-manual_1212080048.webp)
4
Recording an ECG
4.1
Placing the Electrodes
Page 48
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
4.1.3
Standard 12-lead
C1 Red
R Red
N Black
C2 Yellow
C3 Green
C4 Brown
C6 Violet
C5 Black
F Green
L Yellow
C1 white / red
V1 brown / red
Fourth intercostal space at the right sternal border
C2 white / yellow
V2 brown / yellow
Fourth intercostal space at the left sternal border
C3 white / green
V3 brown / green
Midway between C2 and C4
C4 white / brown
V4 brown / blue
Left mid-clavicular line in the fifth intercostal space
C5 white / black
V5 brown / orange
Left anterior axillary line on the same horizontal level as C4
C6 white / violet
V6 brown / violet
Left mid-axillary line on the same horizontal level as C4
L yellow
LA black
Left arm
R red
RA white
Right arm
F green
LL red
Left foot
N black
RL green
Right foot
• Auto Interpretation is only generated when standard 12-lead electrode lead config-
uration is set.
• The lead configuration is set in the lead configuration screen selected at the start
of the recording.
• The electrode resistance is constantly monitored in the recording screen and a
lead-off indication displayed if the resistance is too high.
• When making an ECG with a child it is sometimes physically difficult to position all
electrodes. When this is the case electrode C4 can be placed on the right side of
the chest.
IEC Label
AHA Label
Electrode Placement
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...