7
Exercise ECG Recording
7.5
During the Test
Page 78
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
7.5.3
Trend Graphs
The trend graphs at the top of the screen give the following information:
• HR, BP, METS and Load Trend over the test.
• ST Amplitude and slope
• Reference ST amplitude / actual ST amplitude of every lead
– The J-point is defined in the reference complex.
7.5.4
Blood Pressure
Automatic BP Measurement During the Test
Automatic BP measurements can be taken for any selected stage as set in the
protocol (
see para. 11.14.7, Protocol Editor, page 135)
. These can either be entered
manually (measurement flashes when a measurement needs to be taken), or a
measurement is taken automatically by a BP device attached to the unit. When BP
measurements are defined in the protocol, BP measurement is started 10s after the
pre phase, 50s before the end of work/recovery step or instantly if work/recovery
stage is shorter than 50s.
A BP measurement can also be taken at any time during the test as follows:
Initiating a BP Measurement from an Attached Device
Measurement in progress
Initiate BP
Measurement
When an NIBP device has been defined (
see para. 11.14, Exercise ECG, page 133)
blood pressure measurements are initiated by clicking on the BP icon
. During
the measurement the measurment field displays a timing indicator. On completion,
the BP measurement is displayed.
Manually Entering a BP Measurement
Click to manually enter
BP Measurement
• Click on the measurement - the BP entry screen is displayed.
Enter blood pressure
Systolic:
Diastolic:
OK
Cancel
7.5.5
SpO
2
• SpO
2
measurements are taken either entered manually or taken by an SpO
2
de-
vice attached to the unit or incorporated in an BP unit (for example the SCHILLER
BP-200 with SpO
2
option).
Manually entering an SpO
2
Measurement
SpO
2
measurements are entered manually by clicking on the SpO
2
icon.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...