Safety Notes
1
User Guide
Extra Precautions for Spirometry
1.14
A
rt.
no.
:
2.
511
335
Rev.
a
CARDIOVIT CS-104
1.14 Extra Precautions for Spirometry
In order to obtain correct predicted values and diagnosis, it is important that all
patient data is entered correctly. In particular gender, date of birth, ethnic, height
and weight must be entered.
The unit must be calibrated before the first pulmonary function test of the day and
after every significant temperature change.
False measurements can result when the sensor is not held vertically - ensure
that the sensor is held upright at all times.
The disposable mouthpiece of the
SpiroScout sensor
is designed for one-time
use to eliminate the danger of cross contamination - do not use the mouthpiece
for more than one patient
. Do not attempt to clean the mouthpiece.
See the CARDIOVIT CS-104 Spiro User Guide for full safety precautions.
1.15 Electrical and Power Source Related
To avoid the risk of electric shock, this equipment must only be connected to a
mains supply with protective earth.
Operating the device without the correctly rated fuse, or with defective cables,
constitutes a danger to life. Therefore:
– Do not operate the unit if the earth connection is suspect or if the mains lead
is damaged or suspected of being damaged.
– Damaged cable connections and connectors must be replaced immediately.
– The electrical safety devices, such as fuses, must not be altered.
Ruptured fuses must only be replaced with the same type and rating as the orig-
inal.
Only use power cords provided by the dealer to ensure safety and EMC compli-
ance.
The CARDIOVIT CS-104 unit must be connected to an approved power source
as shown on the specification label.
The power cords must not be damaged. Applied pressure, heat, and stress can
damage the power cord.
The power cords must be routed properly so as to help prevent people from step-
ping on the cords or the cords being run over by, for example, the trolley
Do not overload the mains outlet or extension cords. Electrical shocks or fires
may occur from overloading.
Do not touch the power source during a thunderstorm.
If your hands are wet, do not touch the plug.
Do not pull the power cord to remove it from the mains socket because this can
damage the cable. Use your thumb and index finger to grip the plug itself.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...