11
System Settings
11.14 Exercise ECG
Page 134
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
Max. load calculation
Select between:
•
Max step load
- The current load within the work phase.
•
Time
- The current load of the step is taken as the max. load if the elapsed time
within the step is greater / equal than the configured x second. If the load step is
less than x, the load of the previous step is used as the max. reached load. When
this option is selected the field below becomes active to enter the number of sec-
onds.
•
Interpolated
- Max load = SL1 +
* t2
• SL1: Load of previous step
• SL2: Target/set load of current step
• t1: Step duration in seconds
• t2: Elapsed step time in seconds
SL2 – SL1
t1
Step timer
Display
remaining time
or elapsed time
Recovery
Select to recover either
Direct
or initialise recovery. This determines the action and
recovery icon text to end the test. Direct means that the recovery phase is entered
directly; initialise recovery means that the current exercise stage is completed before
recovery stage is entered (set by user).
Average complex
Check this box to show the average reference complex (
.
Show chart legend
Check this box to display the legend for the three graphs in the upper area (can be
enabled / disabled (
see para. 7.5, During the Test, page 75)
.
11.14.3
Scale
Set the default amplitude and speed scale for the following:
• Rhythm
• Continuous - (multi-line single lead view)
• Average
11.14.4
Events
Define the wording for manual events. As many entries as required can be added
separated by a carriage return (that is new line). Each line is handled as a single event
entry is available during the recording when the Manual event button is clicked (
.
11.14.5
Ergo Devices
Ergo Device
Select the ergo device that wilt be used for the test - bike or treadmill.
Bicycle type / Treadmill type / NIBP
device
Select the ergometer type / NIBP device from the list. If not listed select unsupported.
Bicycle port/ Treadmill port / NIBP
device
Select the COM port to which the ergometer / NIBP device is connected.
When a NIBP device is defined that supports SpO
2
measurement, an extra box is
displayed in the NIBP device settings to Enable SpO
2
trend.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...