11
System Settings
11.15 Spirometry
Page 136
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
4. Define the new protocol name and click Ok.
– For bike protocols an extra check-box is given when a new protocol is set -
Calculate load as Watt per Kilogram
. When checked the work and load
stages in the protocol change from Watt, to Weight - Load (W/kg) and the load
applied during the exercise steps are calculated as a product of the patients
weight:
Duration
Weight-Load (W/kg)
NIBP
Exercise 16:00 min
0.35
0.7
1.0
01:30
01:30
01:30
5. Check the ramp box for a ramp protocol (step increase is carried out over the step
and not a ‘step increase’).
6. Check the warmup step to define a warmup stage.
7. For each step check the BP box to initiate a BP measurement for that stage or
display the BP input screen for manual entry.
Note:
– A BP measurement is taken 10s after the pre phase and after the warmup
phase (when a warmup phase has been set in the protocol).
– A BP measurement is initiated when checked for a stage, as follows:
– 50s before the end of work/recovery step OR
– instantly if work/recovery stage duration is less than 50s.
8. Define work and recovery stages load or elevation / speed.
9. Define the protocol for inclusion in the Protocol option when taking a test (see
paragraph above).
Auto Protocol Editor
When the Auto protocol editor is selected
, the protocol can be overwritten with a
set number of work steps and recovery steps, and linear load increase and step
duration.
Replace all steps
Work
Generate
Cancel
Recover
No. of steps Initial load(W) Load delta (W) Step duration (min)
11.15 Spirometry
See CARDIOVIT CS-104 Spiro User Guide.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...