![Schiller CARDIOVIT CS-104 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/schiller/cardiovit-cs-104/cardiovit-cs-104_user-manual_1212080055.webp)
Page 55
Recording an ECG
4
User Guide
Placing the Electrodes
4.1
A
rt.
no.
:
2.
511
335
Rev.
a
CARDIOVIT CS-104
4.1.10
Right Precordials (C3r-C6r)
Since the treatment of an infarction might depend on the influence of the right
ventricle, it is suggested to perform additional recordings with right precordial leads in
the case of an acute infarction of the right ventricle's inferior wall (Circulation 2007).
C1 Red
R Red
N Black
C2 Yellow
C3r Green
C4r Brown
C6r Violet
C5r Black
F Green
L Yellow
C1 white / red
V1 brown / red
Fourth intercostal space at the right sternal border.
C2 white / yellow
V2 brown / yellow
Fourth intercostal space at the left sternal border.
C3r white / green
V3 brown / green
Designated point halfway between C1 and C4r.
C4r white / brown
V4 brown / blue
Fifth intercostal space on the mid-clavicular line.
C5r white / black
V5 brown / orange
Anterior axillary line on the same horizontal level as C4r.
C6r white / violet
V6 brown / violet
Mid-axillary line on the same horizontal level as C4r.
L Yellow
LA Black
Left arm
R Red
RA White
Right arm
F Green
LL Red
Left foot
N Black
RL Green
Right foot
.
IEC label
AHA label
Positioning
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...