![Schiller CARDIOVIT CS-104 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/schiller/cardiovit-cs-104/cardiovit-cs-104_user-manual_1212080014.webp)
1
Safety Notes
1.6
Contraindications
Page 14
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
1.6 Contraindications
The CARDIOVIT CS-104 is not designed:
– for sterile use
– for use in areas where there is any danger of explosion or in the presence of
flammable gases such as anaesthetic agents
– for direct cardiac application
– for use in an MRI suite
– for outdoor use.
1.7 Responsibility of the User
The CARDIOVIT CS-104 with the MS-12 ECG Recorder or SpiroScout must only
be used by qualified physicians or trained medical personnel.
The numerical and graphical results and any interpretation given must be
examined with respect to the overall clinical condition of the patient and the
general recorded data quality.
Ensure that personnel have read and understood these operating instructions. In
particular this section safety notes must be read and understood.
The operator is responsible for compliance with all applicable accident prevention
regulations and safety regulations.
The safety, reliability and performance of the device can only be guaranteed
when the maintenance intervals as stated in the maintenance section are
observed.
1.8 Organisational Measures
Observe the operating and maintenance instructions. Keep all instructions in an
accessible place for reference when required. Make sure that they are always
complete and legible.
In addition to this user guide, legal and other binding regulations for the
prevention of accidents and for environment protection must be observed
1.9 Safety Facilities
Operating the device without the correctly rated fuse or with defective cables
constitutes a danger to life! Therefore:
–Do not operate the unit if the earth connection is suspect or if the mains lead, the
power supply unit or the device is damaged or suspected of being damaged.
–Damaged cable connections and connectors must be replaced immediately.
–Electrical safety devices, such as fuses, must not be modified.
–Fuses must only be replaced with the same type and rating as the original.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...