Safety Notes
1
User Guide
Safety-Conscious Operation
1.10
A
rt.
no.
:
2.
511
335
Rev.
a
CARDIOVIT CS-104
1.10 Safety-Conscious Operation
©
The CARDIOVIT CS-104 is CF classified. It is defibrillation protected only
when the SCHILLER original patient cable is used. As a precaution however,
when possible remove the electrodes before defibrillation.
Ensure that the patient is informed about the procedure for stress testing and is
aware of the risks (for example, of falling on a running treadmill). Ensure the
patient is aware of the location and use of the emergency stop knob when using
a treadmill.
Do not touch the unit casing during defibrillation.
To ensure patient safety, none of the electrodes, including the neutral electrode,
nor the patient or any person with simultaneous patient contact, must come in
contact with conductive parts, even when these are earthed.
There is no danger when using the device for a patient with a pacemaker fitted.
However, data transmission modules could affect the pacemaker functionality.
To prevent a pacemaker malfunction, a distance of at least 25 cm must be kept
between the device and the pacemaker as soon as the Bluetooth module is
activated.
Do not place any liquids on the unit.
Only use the original SCHILLER patient cable and only use accessories and
disposables recommended or supplied by SCHILLER. Use of other than
recommended or supplied parts may result in injury, inaccurate information and/
or damage to the unit.
If the patient cable should become defective after defibrillation, an electrode
becomes displaced, or an electrode resistance is too high, a lead-off indication is
displayed.
Precautions for Bluetooth pairing:
–ensure that no two sensor pairing processes are started simultaneously to pre-
vent incorrect pairing, and
– ensure that only one recorder is in range of the receiver during advertising/
pairing.
1.11 Trolley Installations - Transporting &
Placement
Be cautious when moving the Trolley. Quick stops, excessive forces and uneven
surfaces may cause the trolley to overturn thus risking the unit to fall to the
ground.
If the unit falls to the ground, turn the power off immediately and disconnect from
the mains. Contact a SCHILLER approved service centre. Continual use of the
unit can result in fire or electric shock.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...