![Schiller CARDIOVIT CS-104 Скачать руководство пользователя страница 150](http://html1.mh-extra.com/html/schiller/cardiovit-cs-104/cardiovit-cs-104_user-manual_1212080150.webp)
14
Cleaning and Disinfection
14.4 Cleaning Cable Assemblies
Page 150
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
14.4 Cleaning Cable Assemblies
1. Before cleaning, inspect the cable for damage as detailed above.
2. Wipe the equipment with a cloth slightly moistened (not wet) with one of the
approved solutions listed below.
3. Gently grip the cable with the damp cloth in the center of the cable and slide the
cable through the cloth 20 cm at a time until clean. Do not clean the whole length
in one single action as this may cause bunching of the insulation sheathing.
4. Wipe off any excess cleaning solution. Do not let the cleaning solution run into or
accumulate in connector openings, latches, or crevices. If liquid gets into
connectors, dry the area with warm air.
14.5 Cleaning Solutions
14.5.1
Construction Material
Ensure that all cleaning materials used on the MS-12 USB ECG Recorder and cable
assemblies are suitable for the material used in its construction. The list of cleaning
solutions and disinfectants are provided as a general guide. If in doubt about the
suitability of a cleaning solution or disinfectant, check that the solution is suitable for
the materials as follows:
14.5.2
Approved Cleaning Solutions
• 50% solution isopropyl alcohol
• Neutral mild detergent solution
• All products designed for cleaning plastic.
MS-12 ECG Recorder
Section
Material
Housing
PC/ ABS (polycarbonate-
ABS)
Patient Cable
PE (Polyethylene)
USB Cable
PE (Polyethylene)
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...