Trouble Shooting
16.1 Error Messages
Page 156
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
16 Trouble Shooting
The following table gives common faults, causes and possible remedies.
Interferences, lines on the
display
• Excessive EMC interferenc-
es
Check for sources of excessive EMC interfer-
ences.
‘Noisy’ traces.
• High resistance electrode
contact.
• Patient not relaxed.
• Incorrect filter settings.
Check electrode contact
ity Indication on the ECG Trace, page 61)
Re-apply electrodes.
Ensure that the patient is relaxed and warm.
Activate filter
(see para. 4.8, Filter, page 63)
Ensure mains filter is correct for mains supply
(
see para. 11.11.2, Power Line, page 132)
If the trace is still noisy call your local SCHILLER
representative.
Control icons, buttons or
options not displayed
• Options not licensed,
• Configuration does not sup-
port option
• Control /action /view but-
tons not defined for view
Check licensing options (
Define for view (
see para. 3.8, Display Configu-
16.1 Error Messages
System messages, error messages, and fault messages are provided in the program
in the event of an error or if an action is attempted that is not licensed or authorised
(by user privilege). Where possible suggested action(s) to be taken to resolve the
problem are indicated in the message.
Fault
Possible Causes and indicators
Remedies
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...