3
- 3
3
Measuring The Water Sample
(5) When the zero-check is completed, add the reagent into the residual chlorine meter container, then mix well for
approximately 20 seconds.
Add the total chlorine reagent that was specified by the
manufacturer of the residual chlorine concentration meter.
(6) Place the container into the residual chlorine meter again and measure the concentration. The measured value
multiplied by 10 is the hypochlorous acid concentration.
The displayed value is the
measurement value.
(7) The hypochlorous acid concentration shows normal values if it is within the designated range of 2 – 5 ppm.
(8) If the hypochlorous acid concentration is not within the designated range (2 – 5 ppm), change the electrolysis
condition according to the table below and replace the water in the electrolyzed water tray. Then measure the
concentration again.
I
Judging the hypochlorous acid concentration
· Standard range: over 2 – 5 ppm
· If the measured concentration is not within the standard range (2 – 5 ppm), change the electrolysis condition
according to the table below.
Measured concentration < 2 ppm
5 ppm < measured concentration
Old
New Old
New
setpoint
setpoint
setpoint
setpoint
No. 1
No. 2
No. 2
No. 1
No. 2
No. 3
No. 3
No. 2
No. 3
No. 4
No. 4
No. 3
No. 4
No. 5
No. 5
No. 4
No. 5
No. 7 *1
No. 6
No. 4
No. 6
No. 7 *1
No. 7
No. 6
*1: If setting this point, dissolve 1 g of salt in the water tank before inserting the tank into the unit.
07-032 VW-VF10BG_TD 2/15/07 11:24 AM Page 4