- 80 -
㻌
㻌
z
Removal of MCO-HP on the way of decontamination
# Operation
Key
Display/Remark
1
Turn the unit off and on.
Power switch
“H2O2 Decontamination”
2 When ‘Warming up’ is displayed, press
MENU key and select ‘about’ and press
ENTER key.
MENU
ENTER
To avoid the device turning back
to resolution process, ensure to
press “MENU” key and select
‘about’ and press “ENTER” key
when ‘Warming up’ is displayed.
3
Remove MCO-HP and press MENU key
and ENTER key.
MENU
ENTER
“Top Screen”
4
Ensure electric lock should be unlocked.
Ensure outer door should be
opened.
(3) Automatic CO
2
Density Calibration Add-on Kit (MCO-SG)
z
Setting of Standard CO
2
and O
2
density
# Operation
Key Display
1
-----
“Top
Screen”
2
Press MENU key.
MENU
“MENU” window
3
Select Tool and press ENTER key.
ENTER
“Select Tool”
4
Select “Standard Gas setting” on page 2.
Press ENTER key.
ENTER
“Standard Gas Setting” screen
5 Input standard gas density by numeric
keys.
Numeric key
6
Press MENU key.
MENU
“MENU” window
7
Select ‘OK’ and press ENTER key.
ENTER
z
CO
2
and O
2
density calibration
# Operation
Key
Display/Remark
1
-----
“Top
Screen”
2
Press MENU key.
MENU
“MENU” window
3
Select Tool and press ENTER key.
ENTER
“Select Tool”
4
Select “Standard Gas calibration” on the
2
nd
page of “Select Tool” screen.
Press ENTER key.
ENTER
“Standard Gas Calibration” screen
(System check starts)
After passing system check,
confirmation screen is shown to
start calibration.
5
Select ‘OK’ in “MENU” window and press
ENTER key to start CO
2
calibration.
ENTER
6
After
CO
2
calibration is complete,
“Top Screen” is shown.
Содержание MCO-19M
Страница 7: ... 4 Dimensions Power cord ...
Страница 16: ...Wiring diagram 13 ...
Страница 17: ...Circuit diagram Main PCB 14 ...
Страница 18: ... LCD PCB 15 ...
Страница 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Страница 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Страница 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Страница 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...