BASIC PARAMETERS
Setting the chamber temperature, CO
2
and O
2
density
The setting procedure for the chamber temperature, CO
2
and O
2
density are given below. (Default
settings:
Chamber temperature: 37
o
C, CO
2
density: 0%, O
2
density: 20%)
The Incubator automatically starts operation using these settings after the power is turned ON.
1.
From the Top Display, press the MENU Key to access the menu. Select
Set
and press the ENTER
Key.
2.
The Stand-By Setting Display will be displayed. Set the parameters.
The setting ranges of the parameters are as follows:
Ɣ
Temperature: 0 to 50
o
C. To 37.0
o
C, enter 370.
Ɣ
CO
2
density: 0 to 20%. To set 5.0%, enter 050.
Ɣ
O
2
density: 1 to 18 and 22 to 80%. To set 5.0%, enter 050.
3.
Press the MENU Key to display the menu after setting the parameters. Select
OK
and press the
ENTER Key. The settings will be saved.
Note:
When starting the Incubator for the first time or after not using it for an extended period of time, allow at
least 4 hours for the chamber temperature, humidity, CO
2
sensor and O
2
sensor to become stable after
setting the desired chamber temperature at a 0% CO
2
density and a 20% O
2
density. Then change the
setting to the desired CO
2
and O
2
density.
5
.
1
T e m p
3 7 . 0
o
C C
O
2
5 . 0 %
O 2 5 . 0
%
2
3
4
5
S
t a t u s
6
O K
7
37
.
0
5
.
0
5
.
0
M E N U
S e t
L
o
g
T
o
o
l s
1
S t a n d - b y S e t t i n g
2
T e m p e r a t u r e
3 7 . 0
o
C ( 0 . 0
o
C - 5 0 . 0
o
C )
3
C
O
2 D
e
n s i t y 0 . 0 %
( 0 . 0 % - 2 0 . 0
%
)
4
O
2 D
e
n s i t y 2 0 . 0 %
( 1 . 0 % - 1 8 . 0
%
)
5
( 2 2 . 0 % - 8 0 . 0 % )
K e y L o c k
0 ( 0 . U n l o c k 1 . L o c k )
6
H i g h L i m i t 5 2 . 0
o
C
M E
N
U
O K
C a n c e l
- 132 -
25
Содержание MCO-19M
Страница 7: ... 4 Dimensions Power cord ...
Страница 16: ...Wiring diagram 13 ...
Страница 17: ...Circuit diagram Main PCB 14 ...
Страница 18: ... LCD PCB 15 ...
Страница 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Страница 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Страница 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Страница 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...