Tel: 886.909 602 109 Email: [email protected]
42
c) Reduce the resolution bandwidth until you can see the distortion product:
Press [BW] and reduce the resolution bandwidth with the step key [↓].
d) Adjust both signal generators until the amplitudes of input signals are the same.
Press [Maker], [Delta], [Peak] and [Next Peak]. Adjust the signal generator corresponding to the marker until the
amplitude difference is zero. If required, reduce the video bandwidth.
Fig. 4-6 Signals at Center of Display of Spectrum Analyzer
e) Set the reference level and make the signal peak at the reference level:
Press [Peak] and [Peak Search], and read the peak power.
Press [Amplitude], [Reference Level].
In order to achieve the best measurement accuracy, the signal peak of the signal generators should be set at the
reference level, as shown in Fig. 4-7.
Содержание S3302SB
Страница 1: ...S3302 Series Handheld Spectrum Analyzer User Manual Saluki Technology Inc...
Страница 136: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 136 Fig 7 1 FM Demodulation Analysis Results...
Страница 138: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 138 Fig 7 3 AM FM PM Analyzer Menu continued...
Страница 189: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 189 Fig 11 3 Schematic diagram of map location...