background image

Tel: 886.909 602 109 Email: [email protected]

www.salukitec.com

26

[Mode]:

Set the measurement mode, including spectrum analysis, interference analysis, analog

demodulation, power measurement, channel scanning and field strength measurement.

Reset button

Press [Reset] to shut down the instrument and power up it again.

4

Reset button

Press [Preset] to shut down the instrument and power up it again.

5

Power switch

Start and shut down the spectrum analyzer. When the instrument powered up with the external power supply
through the adapter is in the “Standby” state, the yellow indicator near the power switch will be ON. If the power
switch is pressed for more than 3s, the indicator will turn green, indicating that this instrument is “operating”. When
the power switch is pressed for more than 3s in the operating state, the spectrum analyzer will be shut down.
Note: “Flicker” of the indicator indicates that the electricity quantity of the battery is not full at present and the battery
is being charged.

Содержание S3302SB

Страница 1: ...S3302 Series Handheld Spectrum Analyzer User Manual Saluki Technology Inc...

Страница 2: ...ctrum analyzer 9kHz 26 5GHz S3302C handheld spectrum analyzer 9kHz 32GHz S3302D handheld spectrum analyzer 9kHz 44GHz Standard pack and accessories No Item 1 Main Machine 2 Power cord 3 Power adapter...

Страница 3: ...1 S3302 16 Interference analyzer option Waterfall RSSI S3302 17 AM FM PM analyzer option To Realize Modulation Characteristics Analysis of AM FM PM Signals S3302 18 Channel scanner option To Realize S...

Страница 4: ...ly S3302 37 Tracking generator 100kHz 9GHz For S3302SC only S3302 38 Orientation analysis option Built In software need option10 50 and directional Antenna S3302 39 Coverage map option Built In softwa...

Страница 5: ...140MHz 430MHz S3302 54 Passive directional antenna 700MHz 4GHz Passive log periodic antenna S3302 55 Passive directional antenna 700MHz 6GHz Passive log periodic antenna S3302 56 Passive directional...

Страница 6: ...in this document rests with Saluki Saluki Tech owns the copyright of this document which should not be modified or tampered by any organization or individual or reproduced or transmitted for the purp...

Страница 7: ...886 909 602 109 Email sales salukitec com www salukitec com 7 Contacts Service Tel 886 909 602 109 Website www salukitec com Email sales salukitec com Address No 367 Fuxing N Road Taipei 105 Taiwan R...

Страница 8: ...on 2 Structure of Spectrum Analysis Menu 70 Section 3 Description of Spectrum Analysis Menu 76 Chapter V Interference Analyzer Measurement Mode optional 109 Section 1 Introduction to Typical Measureme...

Страница 9: ...h Menu 160 Section 3 Description of Field Strength Menu 162 Chapter X Signal Analyzer Mode optional 169 Section 1 Introduction to Typical Measurements 169 Section 2 Structure of Signal Analysis Menu 1...

Страница 10: ...tellite communications wireless communications radar surveillance electronic warfare and electronic surveillance precision guided and other industries The following safety notes are used throughout th...

Страница 11: ...ous intelligent measurement functions such as the channel power occupied bandwidth adjacent channel power tune listening emission mask and C N Due to integration of the 8 4 inch LCD and capacitive tou...

Страница 12: ...CD and touch screen integration various display modes automatic adjustment of backlight brightness etc Operating temperature range 10 C to 50 C This instrument can be powered up with the battery or ad...

Страница 13: ...ous kinds of electronic weapon equipment such as the radar communication electronic countermeasures and reconnaissance and precision guidance On site testing and diagnosis of transmitter and receiver...

Страница 14: ...ing list 1 Please check the articles above against the order contract and packing list If you have any questions please contact us in line with the contact information in Preface or contact the busine...

Страница 15: ...the battery storage temperature range is 20 C to 60 C the battery must not work continuously in a long time at high temperature so as to avoid risks arising from high temperature It is recommended to...

Страница 16: ...0A DC power supply Voltage 15V Current 3A min Power supply with built in battery S3302 series instruments can be powered up with the rechargeable lithium ion battery The battery will discharge if it...

Страница 17: ...cable When wearing a anti static wrist band hold on to the casing of the cable connector and connect the casing to the other end of the cable before removing the short circuiter 5 Input output port p...

Страница 18: ...n order to facilitate the long time field test and prevent the test interruption caused by insufficient battery capacity the user may purchase the standby battery It is recommended to purchase one of...

Страница 19: ...he following methods Observe the battery icon on the system status bar to roughly know the battery capacity If only 15 exists in the battery icon replace or charge the battery promptly Take out the ba...

Страница 20: ...ttery power more than 5 should be charged for about 4h in the shutdown status Section 4 User Inspection 1 Start up of spectrum analyzer Connect S3302 series spectrum analyzer to the external power sup...

Страница 21: ...splay zone S3302 series spectrum analyzer is equipped with one 8 4 inch color touch screen Parameter setting and information display can be performed by means of touch thus eliminating the troublesome...

Страница 22: ...ation display zone and bottom function buttons 7 in total 1 Top information display zone The top information display zone in the top of the screen shows the system date time and the current power supp...

Страница 23: ...ery capacity is less than 10 the spectrum analyzer may further operate for about 10min In this case measurement results should be kept promptly If the battery capacity is less than 5 the spectrum anal...

Страница 24: ...includes two kinds of information Local showing the current operating status of the spectrum analyzer local or remote control The current center frequency and span information are displayed at the bot...

Страница 25: ...e the value When the knob is rotated clockwise the value will be increased otherwise the value will be decreased The knob can be used with the UP DOWN key to change the value with the same step as the...

Страница 26: ...wer switch Start and shut down the spectrum analyzer When the instrument powered up with the external power supply through the adapter is in the Standby state the yellow indicator near the power switc...

Страница 27: ...tor is grounded 2 2 Test ports 1 RF input port It is used for inputting the signal to be tested The indicator of the test signal input port of S3302 series is 50 The N type female port is applied for...

Страница 28: ...xternal PC which is used for program control or data transmission of S3302 series through the program control commands or function library Equipment drive should be installed for connecting the device...

Страница 29: ...nstrument symbols warning labels in the figure mean that the maximum power of the test port input is 30dBm and the maximum input DC level is 16VDC When the device is in operation the user is not allow...

Страница 30: ...er limit field strength measurement channel power occupied bandwidth adjacent channel power C N emission mask IQ capture audio monitor etc Signal Track If the drift signal is tested the active marker...

Страница 31: ...ore than the set upper limit or less than the set lower limit the spectrum analyzer will send the alarm signal Field strength measurement The spectrum analyzer has the function of field strength measu...

Страница 32: ...S3302 series is a basic operating mode The Quick Operation Guide of S3302 Series Spectrum Analyzer has introduces some typical measurement of this mode including some basic measurement methods such a...

Страница 33: ...filter circuit test it indicates that the system is faulty The channel power measurement is applied to evaluate the communication transmitter and determine the quality of RF transmission by compariso...

Страница 34: ...s g Set the resolution bandwidth and video bandwidth of the spectrum analyzer Press BW and RBW Auto Man and set the resolution bandwidth as 30kHz Press BW and VBW Auto Man and set the video bandwidth...

Страница 35: ...width measurement function of S3302 series spectrum analyzer to measure the occupied bandwidth of the signal 1 Definition of occupied bandwidth The occupied bandwidth refer to the bandwidth including...

Страница 36: ...e spectrum analyzer into the default state Press Reset c Set the center frequency Press frequency and Center Freq to set the center frequency with number keys Set the center frequency of the spectrum...

Страница 37: ...measurement results g Select the measurement method Press Measure OBW and Method XdB and select the occupied bandwidth measurement method You can set the method as the power percentage or power drop...

Страница 38: ...ent channel It is generally expressed as the ratio of the power within the specified bandwidth under various offsets of the adjacent channel to the total power of the channel The adjacent channel powe...

Страница 39: ...and video bandwidth Press BW and RBW Auto Man and set the resolution bandwidth as 30kHz Press BW and VBW Auto Man and set the video bandwidth as 30kHz or less e Enable the ACPR measurement Press Measu...

Страница 40: ...est Off On to enable the limit test function If the adjacent channel power is beyond the set limits the screen background will become red for indication 4 Third order IM distortion measurement 1 Defin...

Страница 41: ...gnal generator of other frequencies are allowed However the frequency interval must be approximately 1MHz in this example Fig 4 5 Connection of Third order IM Distortion Measurement System Set the out...

Страница 42: ...and Next Peak Adjust the signal generator corresponding to the marker until the amplitude difference is zero If required reduce the video bandwidth Fig 4 6 Signals at Center of Display of Spectrum Ana...

Страница 43: ...ed In this case relative measurement can be performed easily Press Peak to activate one marker Press Maker and Delta to activate the second marker Press Peak and Next Pk Left or Next Pk Right to set t...

Страница 44: ...tortion level The difference between the third order distortion component level and fundamental signal level is twice of that between the fundamental signal level and third order interception point Th...

Страница 45: ...ce holding function the spectrum analyzer can display the maximum peak amplitude and frequency drift of the input signal The max hold function of track may be used to measure the occupied bandwidth of...

Страница 46: ...ld will maintain the maximum response to the input signal g Activate the trace 2 and set it into the continuous clearing and writing mode Press Trace Trace 1 2 3 and Clear Write h Change the output fr...

Страница 47: ...1GHz and power as 10dBm Connect the output of the signal generator to the input port of the spectrum analyzer as shown in Fig 4 1 Enable the ON state of the radio frequency b Set the center frequency...

Страница 48: ...ndwidth is measured by the noise marker function using the 1GHz external signal a Set the output signal of the signal generator Set the frequency of the signal generator as 1GHz and power as 10dBm Con...

Страница 49: ...o that the trace data are averaged in a longer interval so as to reduce the measurement error e Move the marker to 1GHz Press Maker and rotate the knob on the front panel until the noise marker readin...

Страница 50: ...ers and mixers in the system Most transmission units and signal generators have harmonics and their components need to be measured 1 Identification of distortion generated by spectrum analyzer In the...

Страница 51: ...ess Maker and Delta Fig 4 12 Observation of Harmonic Distortion f Increase the attenuator setting to 10dB Press Atten Auto Man and 10 dB Observe the differential marker reading as shown in Fig 4 13 Th...

Страница 52: ...such delta marker reading 2 Quick harmonic measurement In this section harmonic components are measured of a signal with frequency of 1 GHz and power of 10 dBm generated by the signal generator a Set...

Страница 53: ...the resolution Press Bandwidth Video Bandwidth Auto Man to enable auto off Use key to reduce video bandwidth d In order to improve the measurement accuracy set the peak level of the fundamental wave a...

Страница 54: ...Reference Level for Maximum Accuracy e Activate the second marker Press Delta and Next Peak In this case the fixed marker is on the fundamental wave while the mobile marker is on the peak point of the...

Страница 55: ...ne distortion percentage is to change the unit to volt Move the decimal of the proportion indicated by the differential marker rights for two places to obtain the distortion percentage The minimum rat...

Страница 56: ...enable auto off Use key to reduce video bandwidth d Reduce the span by the signal tracking function Press Peak to activate the signal peak of marker search Press frequency and Signal Track Off On Pres...

Страница 57: ...ping operation Press Peak and Peak Search and read the peak power Press Amplitude and Ref Level and set it as the peak power Adjust the harmonic peak to the reference level The second harmonic amplitu...

Страница 58: ...s to a RF pulse string of the same repetition frequency and constant pulse width shape and amplitude This section introduces several methods of measuring pulse RF signal parameters including how to me...

Страница 59: ...surement results the video bandwidth should be set as 3MHz Press Reset key Press Frequency Center Frequency and 1 GHz Press frequency Span 10 MHz sweep Sweep Time Auto Man and 60 ms Press BW RBW Auto...

Страница 60: ...e Display in Solid Line Form c Read the center frequency of the pulse and amplitude of the main lobe Press Peak The marker reading is the center frequency of the pulse and amplitude of the main lobe d...

Страница 61: ...ta Peak Next Pk Right and Next Pk Right In this case the reciprocal of the frequency difference indicated by the differential marker is the pulse width as shown in Fig 4 23 To obtain the most accurate...

Страница 62: ...r as 20dBm Connect the output of the signal generator to the input port of the spectrum analyzer Set the repetition frequency of pulse modulation as 1kHz and pulse width as 900ns Enable the pulse modu...

Страница 63: ...epetition interval PRI and its reciprocal is the pulse repetition frequency PRF Fig 4 24 Measurement of Pulse Repetition Frequency 4 Peak pulse power measurement Now we have obtain the main lobe ampli...

Страница 64: ...e mode is point frequency source mode to realize the signal output at single frequency point and fixed power and the generator output mode needs to coordinate with spectral analysis function Under the...

Страница 65: ...de frequency characteristics of measured piece by using the synchronous frequency scan of signal source and spectral analysis For details please refer to the following setting procedures a Press Freq...

Страница 66: ...eq to set the start frequency as 2 1GHz b Press Freq Stop Freq to set the stop frequency as 2 6GHz c Press Freq Generator Generator Off On to start the source switch d Press Measure Generator Mode CW...

Страница 67: ...e map can be saved to the instrument for later recall 1 RSSI measurement Main operation steps of RSSI measurement a Press Measure Coverage Map Coverage Map Off On in the default the measurement mode i...

Страница 68: ...28 as below Fig 4 28 Schematic diagram of RSSI measurement 2 Adjacent channel power ratio measurement Main operation steps of adjacent channel power ratio measurement a Press Measure Coverage Map Cove...

Страница 69: ...terval or how to move the distance h Press Measure Coverage Map Distance Time Repeat Time set the time which takes to mark data once on the map i Press Measure Coverage Map Distance Time Repeat Dist s...

Страница 70: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 70 Section 2 Structure of Spectrum Analysis Menu Fig 4 30 Overall Block Diagram of Spectrum Analysis Menu...

Страница 71: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 71 Fig 4 31 Overall Block Diagram of Spectrum Analysis Menu continued...

Страница 72: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 72 Fig 4 32 Overall Block Diagram of Spectrum Analysis Menu continued...

Страница 73: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 73 Fig 4 33 Overall Block Diagram of Spectrum Analysis Menu continued...

Страница 74: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 74 Fig 4 34 Overall Block Diagram of Spectrum Analysis Menu continued...

Страница 75: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 75 Fig 4 35 Overall Block Diagram of Spectrum Analysis Menu continued...

Страница 76: ...e stop frequency with the or key or knob Step frequency Press Frequency Step Freq and set it with number keys on the front panel Select GHz MHz kHz or Hz in the frequency unit menu or set the step fre...

Страница 77: ...to the instrument mode For S3302G series spectrum analyzer the full span is 44 1GHz Zero span Press frequency Span Zero Span and set the span of the current measurement mode as the minimum span Under...

Страница 78: ...ied in the zero span mode IF Out Off On Press Frequency Span IF Out IF Out Off On and enable or disable IF output with keys IF Select 3IF 4IF Press Frequency Span IF Out IF Select 3IF 4IF and select t...

Страница 79: ...on range is 0dB to 50dB Scale Division Press Amplitude Scale Div Adjust the size of the vertical coordinate division on the screen and set the value with number keys on the front panel Select the freq...

Страница 80: ...The Sample mode of the detector will be enabled by this function At the same time traces will be continuously averaged to achieve the smoothing effects SPAN RBW Press BW SPAN RBW to set the ratio of...

Страница 81: ...ctor is selected in the Max Hold mode Neg Peak Enable the negative peak mode In this mode the negative peak level will be displayed on the trace This is applied in self inspection of the broadband MMW...

Страница 82: ...noise marker If the ON state is selected the noise marker will be active Read the noise power at which noise is normalized to 1Hz bandwidth near the active marker In this case the Sample mode of the...

Страница 83: ...r Marker CF Step and set the step of the center frequency The step is equal to the marker frequency If the differential marker function is activated the frequency step is equal to the frequency of the...

Страница 84: ...ind next peak on the right side of the current marker location Max Search Press Peak Max Search to set one marker at the lowest point of the trace The marker frequency and amplitude will be displayed...

Страница 85: ...yzer to enable the interference analyzer mode Refer to Chapter V Interference Analyzer Mode for details AM FM PM Demodulation Press Mode AM FM PM Analyzer to enable the AM FM PM analyzer mode Refer to...

Страница 86: ...ly associated according to the resolution bandwidth frequency bandwidth and video bandwidth Sweep Cont Single Press Sweep Sweep Type to enable the continuous or single sweeping mode Sweep Once Press S...

Страница 87: ...next sweep or measurement is automatically triggered Video Set the trigger mode as Video Once the input signal exceeds the set video triggering level sweeping will be triggered The trigger level value...

Страница 88: ...mode Max Hold Press Trace Max Hold to hold the maximum value of the points on the selected trace In addition the value will be updated based on the new maximum value detected in each sweeping In the...

Страница 89: ...ht Point and Default Limit to view the frequency and amplitude of the limit point and add or delete the limit point Margin Press Limit Margin to set the offset allowance Alarm Off On Press Limit Alarm...

Страница 90: ...Measure OBW to enable the occupied bandwidth function and recall the related functional menu Refer to OBW menu description for details ACPR Press Measure ACPR to select the adjacent channel power func...

Страница 91: ...enus and corresponding test antennas Field Strength Off On Press Measure Field Strength Field Strength Off On to enable or disable the field strength measurement function Recall Antenna Press Measure...

Страница 92: ...er displayed by the spectrum analyzer within the bandwidth while the channel power span refers to the sweeping frequency range of the spectrum analyzer The channel power span should be more than or eq...

Страница 93: ...y dBc on both sides of the frequency point where the signal peak power is The signal power drop dBc can be set by the user Press Measure OBW to set the power percentage in the percentage method dBc Pr...

Страница 94: ...t the adjacent channel bandwidth with number keys Ch Spacing Press Measure ACPR Ch Spacing and set the channel spacing with number keys Limit Test Off On Press Measure ACPR Limit Test Off On to enable...

Страница 95: ...efault setting is 3MHz Noise BW Press Measure C N Noise BW and set the noise bandwidth with number keys The default setting is 3MHz Frequency Offset Press Measure C N Offset and set the frequency offs...

Страница 96: ...it file to be recalled Reference power Press Measure Emission Mask Ref Power to set the reference power setting mode The peak power or channel power may be used as the reference power Peak Markers Off...

Страница 97: ...ode to enable the single or continuous IQ capture In the single mode the data will be captured once only In the continuous mode the data will be captured once after each sweeping of the stroke and cap...

Страница 98: ...lation Time Press Measure More 1 2 Tune Listen Demod Time to set the listening time Listen Mode Press Measure More 1 2 Tune Listen Listen Mode to set the listening mode The default setting is the inte...

Страница 99: ...set point frequency menu will be infective and the output power power offset frequency offset and transmission measurement will be effective power 0 0 dBm press Measure More 1 of 2 Generator Power 0 0...

Страница 100: ...ctive when the tracking mode and normalized are activated Ref Position 0 press Measure More 1 of 2 Generator Trans Meas Ref Position 0 to activate the normalized and adjust the ref position so as to a...

Страница 101: ...asurement Setup to open the measurement submenu Distance Time press Measure More 1 of 2 Coverage Map Distance Time to open the distance time submenu Start Collection press Measure More 1 of 2 Coverage...

Страница 102: ...a press Measure More 1 of 2 Coverage Map Save Recall Recall Data to recall the saved data RSSI press Measure More 1 of 2 Coverage Map Measurement Setup RSSI to open the RSSI measurement submenu ACPR p...

Страница 103: ...ints on the map Excellent 0 0dBm press Measure More 1 of 2 Coverage Map Measurement Setup RSSI Excellent 0 0dBm to set the color when the measured signal is greater than this value Very good 20 0dBm p...

Страница 104: ...s Measure More 1 of 2 Coverage Map Measurement Setup ACPR Ch Spacing to set the channel interval space Adj Ch Offset press Measure More 1 of 2 Coverage Map Measurement Setup ACPR Adj Ch Offset to set...

Страница 105: ...relevant soft menus such as Head Tail Page Up Page Down Done and Delete You can read the saved data file and recall the corresponding state parameters into current sweeping Save Pic Press Save Recall...

Страница 106: ...hould be paid to the zero frequency signal If the zero frequency signal is above 20dBm LO null calibration should be performed so as to prevent gain compression caused by too large amplitude of the ze...

Страница 107: ...GPS Info and Reset You can enable or disable the GPS function view GPS details and reset GPS Special note The GPS positioning function is optional Internet setting Press System LAN to set the network...

Страница 108: ...custom image version etc Date Format Press System Date Format to set the date format Error Log Press System Error Log and relevant soft menus will pop up including Head Tail Page Up Page Down and Del...

Страница 109: ...cyclic or intermittent signal can be easily observed in the 3D spectrogram display i e frequency amplitude and time The time signal amplitudes are reflected by various colors in the spectrogram displ...

Страница 110: ...the number keys or knob in the vertical direction of the spectrogram The following spectrogram will display the trace information on the line If the time marker value is more than 0 the trace and spe...

Страница 111: ...Sweep Interval to set the sweep interval which represents the sweep time between two adjacent points in each sweeping b Press Auto Save Sweep Time to set the span record time After reaching the set sp...

Страница 112: ...l sales salukitec com www salukitec com 112 The RSSI test structure is shown in Fig 5 2 the displayed contents vary from parameter settings and Fig 5 2 only shows an example Fig 5 2 RSSI Test Diagram...

Страница 113: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 113 Section 2 Structure of Interference Analyzer Menu Fig 5 3 Overall Block Diagram of Interference Analyzer Menu...

Страница 114: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 114 Fig 5 4 Overall Block Diagram of Interference Analyzer Menu continued...

Страница 115: ...th the or key or knob the step should be 1 2 5 the RSSI mode must be set under zero span Starting frequency Press Frequency Start Freq and set it with number keys on the front panel Select the frequen...

Страница 116: ...description for details Special note When the span is changed with the or key or knob the step should be 1 2 5 Full span Press Frequency Span Full Span and set the current span as 44 1GHz Span Press F...

Страница 117: ...set the reference position with the number keys or or key or knob Atten Auto Man Press Ampt Atten Auto Man to enable the auto or manual mode of the attenuator You can change the mode with the number k...

Страница 118: ...ect weak noise signals This instrument supports variable resolution bandwidth settings from 1Hz to 10MHz with the step of 1 3 10 Average Off On Press BW Average Off On The averaging function can be us...

Страница 119: ...re the signal close to the noise floor Neg Peak Press BW Detector Neg Peak This function is used in self inspection of the MMW integrated test instrument in most cases and rarely in tests By using thi...

Страница 120: ...ude will be displayed in dB Noise Marker Off On Press Marker Noise Marker Off On The noise marker will indicate the noise power normalized to 1Hz bandwidth near the active marker In this case the Samp...

Страница 121: ...p and set the step of the center frequency The step is equal to the marker frequency If the differential marker function is activated the frequency step is equal to the frequency of the differential m...

Страница 122: ...Pk Right to find next peak on the right side of the current marker location Max Value Set one marker at the highest point of the trace The frequency and amplitude of the marker will be displayed in t...

Страница 123: ...soft keys When RBW and VBW are changed in the auto mode the sweeping speed will change accordingly The larger the RBW and VBW values are the higher the sweeping speed is and vice versa Based on the co...

Страница 124: ...time will be the whole sweeping period After reaching the span time recording will be stopped Record Off On Press Auto Save Record Off On to enable or disable the Auto Save mode Special note This func...

Страница 125: ...Measure Spectrum to enable the spectrum measurement mode Spectrogram Press Measure Spectrogram to enable the spectrogram measurement mode RSSI Press Measure RSSI to enable the RSSI measurement mode 1...

Страница 126: ...signals up to 40GHz can be tested and the true average power with high dynamic range from 60dBm to 20dBm can be measured The measurement reading will be shown on the display interface of the USB power...

Страница 127: ...mum input power range identified on the USB power probe and ensure that the input is within the specified range so as to prevent the probe from damage as a result of over high power Table 6 1 SAV8723X...

Страница 128: ...he spectrum analyzer The green indicator of the power probe will be ON a moment later c The USB power probe can be shut down after the USB cable is removed In this case the green LED indicator will be...

Страница 129: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 129 Section 2 Structure of Power Meter Menu Fig 6 3 Overall Block Diagram of Power Meter Menu...

Страница 130: ...responding frequency menu or change the frequency with the or key or knob with the step of 10MHz Special note The frequency setting range is related to the selected USB power probe Refer to Schedule C...

Страница 131: ...tion is enabled the current power level will be measured and saved At the same time one power level relative to the saved value will be displayed Offset Off On Press Ampt Offset Off On If the power of...

Страница 132: ...ttention To improve measuring precision of the instrument zero calibration of the instrument is necessary before small signal power measuring of AV8723X Series USB Power Probe Zero calibration refers...

Страница 133: ...er Limit Press Limit Lower Limit Change the lower limit with the number keys or or key or knob with the step of 1dB Alarm Off On Press Limit Alarm Off On The limit alarm should be enabled when the lim...

Страница 134: ...d analyze the carrier power modulation rate carrier offset modulation depth AM modulation frequency offset FM modulation phase deviation PM S N modulation distortion and total harmonic distortion of t...

Страница 135: ...odulation signal so as to accurately demodulate the signal You can observe the bandwidth in the RF spectrum At the same time noise may be produced in the case of too large IF bandwidth which will affe...

Страница 136: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 136 Fig 7 1 FM Demodulation Analysis Results...

Страница 137: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 137 Section 2 Structure of AM FM PM analyzer Menu Fig 7 2 AM FM PM Analyzer Menu...

Страница 138: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 138 Fig 7 3 AM FM PM Analyzer Menu continued...

Страница 139: ...key or knob the step should be 1 2 5 Start freq Press Freq Start Freq and set the starting frequency with the number keys on the front panel Select the corresponding frequency unit or set the starting...

Страница 140: ...or uV in the frequency unit menu or set the value with the or key or knob Special note When the or key or knob is applied the step should be 10dB Reference position Press Ampt Ref Position and set th...

Страница 141: ...of the modulated signal so as to accurately demodulate the signal You can observe the bandwidth of the modulated signal in the RF spectrum Noise may be generated in the case of too large IF bandwidth...

Страница 142: ...er Select Trace and soft menus related to the trace will pop up including RF Spectrum Audio Spectrum and Audio Waveform Refer to the Select Trace menu for details Off Press Marker Off to disable the c...

Страница 143: ...ion Maximum value Set one marker at the highest point of the trace The frequency and amplitude of the marker will be displayed in the right upper corner of the screen The active functions will not cha...

Страница 144: ...t methods percentage or drop dBc In the percentage method the occupied bandwidth of the signal is obtained by calculating the bandwidth of the frequency of certain percentage to the total power of the...

Страница 145: ...reference position of the audio spectrum to facilitate the observation of the audio spectrum line You can change the reference position with the number keys and click OK or with the or key or knob 9 A...

Страница 146: ...to display the audio spectrum only Audio Waveform Press Measure Audio Waveform to display the audio waveform only Summary Press Measure Summary to display the RF spectrum audio spectrum and audio wave...

Страница 147: ...es All operations in this chapter are based on the channel scanner mode which will not be separately described below 1 Channel Scanner Below is an example of the Channel Scanner mode mainly involving...

Страница 148: ...ly involving the following procedures 1 Press Sweep FScan Start Freq and set the center frequency of the starting channel 2 Press Sweep FScan Step Freq and set the frequency step of the measured chann...

Страница 149: ...of Frequency Scanner 3 List scanner Below is an example of the Frequency Scanner mode mainly involving the following procedures 1 Press Sweep MScan Edit List and edit the channel list to be swept The...

Страница 150: ...909 602 109 Email sales salukitec com www salukitec com 150 4 Press Sweep Power Display Curr Max and enable the Current option Set the current power of each channel Fig 8 3 Schematic Diagram of List...

Страница 151: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 151 Section 2 Structure of Channel Scanner Menu Fig 8 1 Channel Scanner Menu...

Страница 152: ...r to List Scanner menu 2 Channel Scanner menu Signal standard Press Sweep Channel Scanner Signal Std and the existing signal standard list will pop up including relevant menus such as Head Tail Page U...

Страница 153: ...Sweep Freq Scanner Bandwidth and set the bandwidth of the measured channel Number of Channels Press Sweep Freq Scanner Number of Channels and set the number of measured channels At most 20 channels ca...

Страница 154: ...odified settings and go back to the List Scanner menu Cancel Press Edit List Cancel to cancel the modified settings and go back to the List Scanner menu 5 Measurement menu Display Graph Table Press Me...

Страница 155: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 155 6 File menu Refer to the description of the file menu in the spectrum analyzer mode...

Страница 156: ...quency and number of points MScan Observe the amplitude and field strength of frequency points in the list by recalling the edited or saved list All the operations in this chapter are based on the Fie...

Страница 157: ...Freq Step Freq and set the step frequency of sweeping 4 Press Freq Points and set the number of sweeping points 5 Press BW BW and set the bandwidth as 150Hz 300Hz 600Hz 1 5kHz 2 4kHz 6kHz 9kHz 15kHz...

Страница 158: ...c diagram of FScan mode is shown in the figure below example Fig 9 2 Schematic Diagram of FScan Mode 3 MScan Main operation steps of MScan 1 Press Edit List Edit List and edit the current list 2 Press...

Страница 159: ...set the dwell time 7 Press Swp Ant Dwell Time Off On to enable or disable the dwell time 8 Press Marker Marker Off On to enable or disable the marker 9 Press Peak to directly set the marker at the max...

Страница 160: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 160 Section 2 Structure of Field Strength Menu Fig 9 4 Overall Block Diagram of Field Strength Menu...

Страница 161: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 161 Fig 9 5 Overall Block Diagram of Field Strength Menu continued...

Страница 162: ...req step Press Freq Freq Step Auto Man and set the point frequency step with the or key or knob The frequency step should be set as 1MHz in the auto mode Freq Track Off On Press Freq Freq Track Off On...

Страница 163: ...front panel Select GHz MHz kHz or Hz in the frequency unit menu or set the starting frequency with the or key or knob Special note The setting range of the starting frequency is 1MHz to 44 1GHz Step...

Страница 164: ...e the attenuation amount of the attenuator will be adjusted automatically according to the reference value In the manual mode the set attenuation amount will always be applied in the attenuator Scale...

Страница 165: ...or 150kHz Detector Press BW Detector to enable the detector menu including Peak Average and Real Time 5 Marker menu Max Search Press Marker Max Search to set the marker at the maximum point Marker Of...

Страница 166: ...tate Dwell Time Off On Press Swp Ant Dwell Time Off On to set the dwell time The dwell time refers to the waiting time at each point in sweeping The default setting is the OFF state and the manual set...

Страница 167: ...enabled the demodulation mode such as FM AM USB and LSB will not be available Volume 95 Press Demod Volume 95 to set the demodulation volume Set the required volume with the number keys or or key or k...

Страница 168: ...rength FScan Press Measure FScan to enable the FScan mode of field strength MScan Press Measure MScan to enable the MScan mode of field strength Special note Ensure that the list exists before enablin...

Страница 169: ...erations of this chapter are based on that the signal analysis mode has been selected This will not be separately described below The main interface of signal analysis measurement mode will measure by...

Страница 170: ...to view the frequency amplitude capture time and other parameters of captured interference signal as shown in Fig 10 2 Fig 10 2 Interface of signal list IQ Capture function can directly save IQ data o...

Страница 171: ...the signal analysis mode the current test task can be saved quickly as an automatically generated file as per current intermediate frequency and test time The storage task of signal analysis is as sho...

Страница 172: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 172 Section 2 Structure of Signal Analysis Menu Fig 10 5 Block diagram of signal analysis menu...

Страница 173: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 173 Fig 10 6 Block diagram of signal analysis menu continued...

Страница 174: ...or key and spinning wheel to change span as per the step size of 1 2 5 Start Freq press Freq Start Freq to use the number key on front panel frequency unit or or key and spinning wheel to set the Star...

Страница 175: ...details please refer to the description of Span Special Notice use or key and spinning wheel to change span as per the step size of 1 2 5 Full Span press Freq Span Full Span to set the current span as...

Страница 176: ...the ref position by using the number key or or and spinning wheel Atten Auto Man press Ampt Atten Auto Man to use the menu to activate the automatic or manual mode of attenuator by number key or or k...

Страница 177: ...improve the ability of detecting weak signal in noise signal This apparatus can change the RBW within 1Hz 10MHz and at the step size of 1 3 10 Video type Lin Log press BW Video type Lin Log to set the...

Страница 178: ...span If amplitude the show value is expressed in dB Marker Noise Off On press Marker Marker Noise Off On to show the noise power which is used to normalize the noise adjacent to the activated marker...

Страница 179: ...Marker Step Freq to set the step size of center frequency namely the frequency step is equal to marker frequency When the difference marker function is activated the frequency step is equal to the fre...

Страница 180: ...Next Pk Right to search the next peak on the right of current marker position Max Search this is used to place a marker at the highest point of trace and display the frequency and amplitude of the ma...

Страница 181: ...sweep time the scan frequency speed will vary with RBW and VBW The bigger RBW and VBW are the quicker the scan speed is the less RBW and VBW are the slower the scan speed is When the minimum sweep ti...

Страница 182: ...iew the historical data Special Notice it can be used only under the waterfall chart measurement mode Restart press Record Restart to restart the scanning 9 Configuration menu Tune Listen press Config...

Страница 183: ...press Config Tune Listen Listen Time to set the listen time Listen Mode press Config Tune Listen Listen Mode to set the listen mode the default state is intermittent listen mode under which the data...

Страница 184: ...ime It should be noted that IQ Capture Time cannot exceed sweep time Capture Type press Config IQ Capture Capture Type to set IQ Capture type as single or continuous Sample Rate press Config IQ Captur...

Страница 185: ...r deactivate the tune listen function IQ Capture Off On press ShortCut IQ Capture Off On to activate or deactivate IQ Capture function Start Capture press ShortCut Start Capture to start IQ Capture Pe...

Страница 186: ...the azimuth measurement of the electronic compass used in the orientation analysis uses the principle of geomagnetism please try to keep a distance from the magnetic object when testing The location a...

Страница 187: ...izontal disc The main procedures of the horizon scan measurement are as follows 1 Press Measure Horizon Scan select horizol scan measurement mode 2 Press Freq Pot Freq set point frequency value 3 Pres...

Страница 188: ...location information of the interference source The main procedures of the map measurement are as follows 6 Press Measure Map Location select map location measurement mode 7 Press Freq Pot Freq set po...

Страница 189: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 189 Fig 11 3 Schematic diagram of map location...

Страница 190: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 190 Section 2 Structure of Location Analyzer Menu Fig 11 4 Block diagram of location analyzer menu...

Страница 191: ...Freq Auto Man to switch to Step Freq Auto Man it is allowed to use the number key or or key and spinning wheel to set the step frequency CF Step Auto Man press Freq CF Step by using number keyto selec...

Страница 192: ...utomatically according to the reference value In the manual mode the set attenuation amount will always be applied in the attenuator Special note The attenuator setting range is 0dB to 50dB step is 10...

Страница 193: ...a Press Swp Ant Recall Antenna and soft menus such as Head Tail Page Up Page Down Done and Delete will pop up You can select and recall the saved antenna factor file Edit Antenna Press Swp Ant Recall...

Страница 194: ...able or disable the audio alarm function Special note If the audio alarm is enabled the demodulation mode such as FM AM USB and LSB will not be available Volume 95 Press Demod Volume 95 to set the dem...

Страница 195: ...t map Zoom In Press Map Zoom in to zoom in the current map scale Zoom Out Press Map Zoom out to zoom out the current map scale Mark Lines Press Map Mark Lines open the mark submenu Mark Current Mark D...

Страница 196: ...nalysis AM FM PM signal demodulation analysis and USB power measurement the RF signal parameters such as the frequency and amplitude can be measured The functional block diagram of S3302 series spectr...

Страница 197: ...rd includes the power conversion and management circuit interface and display circuit FPGA CPLD control and digital signal processing circuit and ARM core module and is used for DC DC conversion and o...

Страница 198: ...s in the case of any fault Please contact us if the failure cannot be removed If S3302 series cannot be started when the Power key is pressed check whether the power supply is normal whether the adapt...

Страница 199: ...nd place it into the packing box together with the tester 2 Use the original packaging materials to pack the instrument properly to minimize damage 3 Set pads in the four corners of the outer packing...

Отзывы: