15 Signal source menu (option)
·[Signal Source Off On]:
press [Measure] → [More 1 of 2]→[Generator]
→[Generator Off On], to activate or deactivate the signal source.
·[Tracking Mode Off On]:
press[Measure]→[More 1 of 2] →[Signal
Source] →[Mode CW Track], to activate or deactivate the signal source
tracking mode.
·[Special Notice]:
the tracking mode switch is effective when the signal
source switch is started. In case of Off, the independent source mode and
the point frequency source output will be activated independent of spectral
analysis. At this time, the set point frequency and output power menus will
be effective but the transmission measurement, power offset and
frequency offset menus are ineffective; in case of Track, the tracking mode
will be activated, the synchronous scanning will be done at signal source
frequency and spectral analysis mode frequency, the set point frequency
menu will be infective, and the output power, power offset, frequency offset
and transmission measurement will be effective.
·[power 0.0 dBm]:
press [Measure]→[More 1 of 2] →[Generator] →
[Power 0.0dBm], to change the output power by using number key or key
[↑] or[↓].
·[Special Notice]:
the output power range is -40dBm~0dBm with step size
of 1dB.
·[CW Freq 1.000GHz]:
press [Measure] →[More 1 of 2] →[Generator]
→[CW Freq 1.000GHz], to change the CW frequency by using number
key or key [↑] or [↓].
·[Power Offset 0.0dB]:
press [Measure]→[More 1 of 2]→[Generator] →
[Power Offset 0.0dB]. When there is gain or loss between generator output
and peripheral equipment, it is allowed to use this parameter to set the
signal source power offset so as to indicate the actual power of system.
This parameter cannot change the actual output power of generator but its
power reading.
·[Special Notice]: The parameter range is -200dB~200dB with default
value of 0dB and step size of 1dB, which is effective when the
tracking mode is activated.
·[Freq Offset 0.000Hz]:
press [Measure] →[More 1 of 2] →[Generator]
→[Freq Offset 0.000Hz], to set the offset value of generator output signal
Содержание S3302SB
Страница 1: ...S3302 Series Handheld Spectrum Analyzer User Manual Saluki Technology Inc...
Страница 136: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 136 Fig 7 1 FM Demodulation Analysis Results...
Страница 138: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 138 Fig 7 3 AM FM PM Analyzer Menu continued...
Страница 189: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 189 Fig 11 3 Schematic diagram of map location...