Tel: 886.909 602 109 Email: [email protected]
117
3 Amplitude menu
· [Reference level]:
Press [Ampt]→ [Ref Level] and set it withe number
keys on the front panel. Select [dBm], [-dBm], [mV] or [uV] in the
frequency unit menu, or set the value with the [↑] or [↓] key or knob.
· [Special note]: When the [↑] or [↓] key or knob is applied, the step
should be 10dB.
· [Reference position]:
Press [Ampt]→ [Ref Position], and set the
reference position with the number keys or [↑] or [↓] key or knob.
· [Atten Auto Man]:
Press [Ampt]→[Atten Auto Man] to enable the auto or
manual mode of the attenuator. You can change the mode with the
number keys or the [↑] or [↓] key or knob.
· [Special note]: The attenuation range is 0dB - 60dB, with the step of
10dB.
· [Scale/Division]:
Press [Ampt]→[Scale/Div] and set it with the number
keys or the [↑] or [↓] key or knob. The scale/division setting range is
0.1dB-20dB.
· [Units]:
The amplitude unit is dBm in the interference analyzer mode.
· [Pre-amplifier Off On]:
Press [Ampt]→[Pre Amp Off On] to turn on or off
the pre-amplifier.