Tel: 886.909 602 109 Email: [email protected]
86
10 Sweep menu
The sweep time refers to the time required for the LO tuning of the spectrum analyzer within the selected frequency
interval, and directly affects the time of one test. It generally changes with the span, resolution bandwidth and video
bandwidth. In the auto mode, the minimum sweep time will be adopted by the spectrum analyzer after
measurement setting. The sweep time can be increased in the manual mode to meet specific measurement needs.
The default sweep mode is linear sweeping, while list sweeping is optional. Multiple segment can be swept
continuously. The functions of editing, saving and recalling the list are available, thus facilitating the operation.
·
[Sweep Time Auto Man]:
Press [Sweep]→[Sweep Time] to adjust the sweep
time of the spectrum analyzer. Use the number keys, step keys or knob to adjust
the sweep time. If “Man” is underlined, the sweep time can be set manually. If
“Auto” is underlined, the sweep time will be automatically associated according to
the resolution bandwidth, frequency bandwidth and video bandwidth.
· [Sweep Cont Single]:
Press [Sweep]→[Sweep Type] to enable the continuous
or single sweeping mode.
· [Sweep Once]:
Press [Sweep]→ [Sweep Once] to sweep once again.
· [Triggering]:
Press [Sweep]→ [Triggering] to select the triggering mode,
including [Free Run], [Video], [External], etc. Refer to the trigger menu for
details.
· [Points [1001]]:
This is optional. Press [Sweep]→[Points] to enable the
soft menu of points select. You can use soft keys such as [201], [501], [1001],
[2001], [4001] to set sweep points.
· [Sweep Mode Lin List]:
Press [Sweep]→[Sweep Mode] to enable the linear or
list mode. In the linear mode, sweeping will be performed based on the linear
frequency interval, and the frequency intervals of adjacent measurement points
are the same. The list mode is optional, in which sweeping will be performed
based on the set frequency range and other parameters in the edited list.
·[Edit List]:
This is optional. Press [Sweep]→[Edit List] to enable the soft menu
of list editing. You can use soft keys such as [Add Seg], [Delete Seg] and [Delete
All] to manage and edit the sweep lit. The selected segment will be in green. After
editing the segment, press [Done] and [OK] to go back to the sweep menu.
Содержание S3302SB
Страница 1: ...S3302 Series Handheld Spectrum Analyzer User Manual Saluki Technology Inc...
Страница 136: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 136 Fig 7 1 FM Demodulation Analysis Results...
Страница 138: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 138 Fig 7 3 AM FM PM Analyzer Menu continued...
Страница 189: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 189 Fig 11 3 Schematic diagram of map location...