R
OUTINE
S
ERVICE
748460-A
6-7
Rosemount Analytical
January 2000
F
IGURE
6-5. L
AMP
A
SSEMBLY
655000
6.2.1 L
AMP
R
EALIGNMENT
After replacing the UV lamp, three items associated with the optical bench must be
adjusted: Lamp, beam alignment, and iris. The procedure for these adjustments
follows:
1. See Figure 6-5. The lamp can be moved vertically and horizontally by adjusting the
clamp rod (item F). The lamp clamp (item G), secures the position of the clamp rod
once alignment is done.
2. See Figure 6-3. Loosen the three screws securing the lamp housing to the
collector housing to allow vertical adjustment of the lamp housing.
3. Power up the analyzer for a minimum of 15 minutes.
4. Set the analyzer to Range 3, and the gain to "1 low."
5. Enter DIAGNOSTICS mode, verify that source current is 300 mA
±
10 mA and all
power voltages are +5, +12, +15 and -15V (each
±
5%).
6. Display the "Sample A" screen (still in DIAGNOSTICS mode). Adjust the horizontal
and vertical positions of the lamp and lamp housing until a maximum reading is
Lamp Housing
654829
1.1
RED
WHITE
BLUE
BLACK
1
2
3
4
LAMP CONNECTOR
"F"
Rod
"A"
Screws
901876
"B"
Screw
809889
Washer
871111
"G"
Clamp
654818
"C"
Lamp Springs (2)
901982
Lamp
654843
"D"
Tube Clamp
654822
"E"
Iris Centerline
Lamp Mount
654825
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......