I
NITIAL
S
TARTUP AND
C
ALIBRATION
Rosemount Analytical
748460-A
January 2000
3-4
To access a function, to store a value in nonvolatile memory or to return to
run mode from span, zero and security screens. The computer
acknowledges ENTER by momentarily flashing [** DATA STORED **] on
the display when used to store a setting in non-volatile memory. Use ENTER
to engage the span and zero functions, which are initiated by the SPAN and
ZERO keys. [CALCULATING SPAN] or [CALCULATING ZERO] will then be
momentarily displayed.
The SHIFT key in conjunction with the ENTER key will return to Run Mode
from any function screen except: 1) the [CALCULATING ZERO/SPAN] screen
and 2) during an auto calibration cycle. SHIFT -
ENTER during operation of zero and span functions will turn off the
appropriate solenoid valve, if connected, for instruments with the Calibration
Gas Control or Auto Zero/Span.
The SHIFT - ENTER combination is the Escape feature.
3.4.3 U
SER
-P
ROGRAMMABLE
K
EYS
Refer to Figure 3-2. F1, F2, SHIFT/F1 and SHIFT/F2 are software-programmable keys
which can be user-programmed to access any frequently used display or sub-menu for
the following modes: Range, Diagnostics, Linearization, Auto Zero/Span, Remote
Range I/O or Alarm, provided the option board selected is still present.
To use this feature, the function keys must be preprogrammed by the user through the
following steps:
1. Access a display or sub-menu that will be frequently used by following the steps in
the particular set of instructions given in the figures in this section until the desired
display is obtained.
2. Press F1, F2, SHIFT/F1 or SHIFT/F2 to program the analyzer to return to this
display from the RUN mode. This will assign F1, F2, SHIFT/F1 or SHIFT/F2 to this
particular display, and will retain those assignments until the key or combination of
keys is reprogrammed using the same procedure described in this section. The
analyzer acknowledges this command by flashing [**KEY SAVED**] on the display.
3. Exit to the RUN mode display by completing the remaining steps in the figure
chosen in Step 1.
4. When the analyzer returns to the RUN mode display, press the key or keys
selected in Step 2 (F1, F2, SHIFT/F1 or SHIFT/F2) to check the setup. The
analyzer will return to the display or sub-menu selected in Step 1.
5. Press SHIFT/ENTER to return to the RUN mode.
SHIFT
ENTER
ENTER
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......