I
NITIAL
S
TARTUP AND
C
ALIBRATION
748460-A
3-17
Rosemount Analytical
January 2000
A
LL
R
ANGE
The “All” range is a special feature that allows the user to use Range 3 coefficients for
all three ranges. It should only be used over dynamic ranges of 3:1 or less. If Range A
is selected in the Linearization Mode, the microprocessor will use the coefficients in
Range 3. Also, if the linearizer is ON and in Range A, the microprocessor will use the
GAIN and TIME CONSTANT from Range 3, regardless of the GAIN and TIME
CONSTANT selected for Ranges 1 and 2.
N
ON
-S
TANDARD
R
ANGES AND
C
OEFFICIENTS
When ordered, special linearization coefficients for non-standard fullscale ranges are
entered in the appropriate range(s) at the factory. If one set of linearization coefficients
has been ordered and a range has not been specified, these coefficients will be for
Range 3.
The user may want the display to read in engineering units (ppm). This response is
linear over the operating range.
The following coefficients will make no correction to the straight line response, but will
cause the analyzer to display engineering units:
A
0
= 0.00000
A
1
= 1.00000
A
2
= 0.00000
A
3
= 0.00000
A
4
= 0.00000
To calculate linearization coefficients other than those installed at the factory, either 11
or 21 data points should be taken. (If urgent, a curve can be created with as few as
4 points. This should be considered temporary and a more accurate curve made with
more points should be created as soon as possible.)
These data points can be obtained with an accurate gas divider or other flow mixing
device. Before calculating coefficients, the data must be normalized to ranges of 0 to 1
units for both % and concentration readings. Then the axis must be reversed as
illustrated in Figures 3-10 and 3-11. A multiple linear regression is then used to
calculate coefficients. (For example: If the range is 0 to 5000 ppm and readings are 0
to 100%, then divide all of the concentrations by 5000 and the readings by 100. Put
the normalized concentrations on the Y-axis and the normalized readings on the
X-axis.
These data points can be entered into any program capable of computing a fourth
order polynomial curve. This curve will be the mirror image of the curve on the
application sheet at the back of the manual, however the linearization coefficients will
be different. Use the coefficients calculated with the curve for linearization coefficients.
Use these coefficients to solve the following equation:
Y = A
0
+ A
1
X + A
2
X
2
+ A
3
X
3
+ A
4
X
4
.
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......