ROUBLESHOOTING
748460-A
5-3
Rosemount Analytical
January 2000
5.5 CASE HEATER
The electronics for the fan and heater are located on the Power Supply Board.
5.6 ERL ERROR MESSAGE
The error message, ERL, indicates one of two events has occurred:
1. An improper linearization coefficient has been inputted, thereby causing a
calculation loop error. This message is only displayed in the Run Mode, and since
the calculation is generally executed in Range Mode or the linearizer setup
screens, the user will not see the ERL message until the analyzer is returned to the
Run Mode.
2. If the user enters a CAL GAS value that is larger than the fullscale value, the ERL
message may be displayed. In such an instance, the software attempts to
calculate a percent-of-fullscale value which is higher than fullscale.
Note:
The ERL message occurs only in extreme cases, i.e., FS value = 5 ppm and CAL
GAS = 5000 ppm. If the user inputs, for example, 50 ppm on a fullscale of 5
ppm, the ERL message will not appear. But when the user attempts to span the
range with this value, the ER# [SPAN POT LMTS] error message will appear
instead.
To correct this condition, do the following:
1. The user should cross-reference the linearizer coefficients entered with those
provided by the factory (or those the user has calculated). Re-enter them properly.
Exit into the Run Mode, and the ERL message should disappear.
2. If the coefficients are correct, then check the fullscale/CAL GAS value relationship
in the Range setup screens. Correct this condition, exit the Range setup screen,
and once the calculation has been automatically redone, the ERL message will
disappear.
Note:
The linearization calculation loop normally takes from 2 to 3 seconds to perform
its task. If either of the two events described above occur, though, causing a
calculation loop error, "WAIT-CALCULATING" may be displayed for up to 20
seconds before it exits with the ERL error message.
Also, when the ERL message is flashing in the Run Mode screen, if the STATUS
key is pressed, the message [EL-LIN.COEFF ERR] will be displayed briefly
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......