6.2 LAMP REPLACEMENT
The ultraviolet lamp is an expendable item and its replacement requires realignment.
Note
Contaminants from oil, dust and fingerprints will reduce the energy and stability
of the lamp. Use clean rubber gloves while handling the lamp. If the lamp
becomes soiled, clean with acetone, rinse with alcohol and then distilled water.
Blow dry with nitrogen.
Acetone is highly flammable and moderately toxic. Use only in a well ventilated,
flame-free, non-smoking location.
Avoid prolonged inhalation of vapors, headache, dizziness and irritation of the
eyes, nose and throat may occur.
Avoid contact with skin, mild irritation may occur. Use rubber gloves and eye
protection when using acetone.
F
IGURE
6-2. S
AMPLE
C
ELL
A
SSEMBLY
Cap, Inlet
654847
O-Ring
899373
Housing
654828
Cell (with covering)
654836
Cap, Outlet
654848
O-Ring
899373
WARNING: FLAMMABLE AND TOXIC CHEMICAL
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......