U
NPACKING AND
I
NSTALLATION
Rosemount Analytical
748460-A
January 2000
2-4
2.5 SAMPLE INLET/OUTLET CONNECTIONS
The standard Model 890 is intended for atmospheric pressure operation only, and
must be vented to either the atmosphere or a collection destination at atmospheric
pressure. Sample inlet and outlet connections are located on the rear panel. All
connections are 1/4-inch ferrule-type compression fittings. See Drawing 654853.
This instrument measures toxic gases. Ensure gas lines are leak-free and
properly vented. Inhalation of toxic gases is highly dangerous and could result
in death.
Also, exhaust gas from this instrument is toxic and equally dangerous. Exhaust
must be connected either to its original source or an appropriate outside vent
using ¼-inch (6mm) tubing minimum.
This analyzer is of a type capable of analysis of sample gases which may be
flammable. If used for analysis of such gases, the instrument must be protected
by a continuous dilution purge system in accordance with Standard ANSI/NFPA
496-1989, Chapter 8.
If explosive gases are introduced into this analyzer, the sample containment
system must be carefully leak-checked upon installation and before initial
startup, during routine maintenance and any time the integrity of the sample
containment system is broken, to ensure the system is in leak-proof condition.
Leak-check instructions are provided in Section 2.8.
Internal leaks resulting from failure to observe these precautions could result in
an explosion causing death, personal injury or property damage.
WARNING: POSSIBLE EXPLOSION HAZARD
WARNING: TOXIC GAS HAZARD
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......