ROUBLESHOOTING
Rosemount Analytical
748460-A
January 2000
5-2
Note
When using the DIAGNOSTICS screen to help make hardware adjustments, set
the analyzer GAIN to X1 Low. This prevents the ADC from being in saturation,
resulting in an erroneous value in the DIAGNOSTICS screen.
5.2 IRIS BALANCE ADJUSTMENT
This mechanical adjustment is provided to balance the system. The adjustment is
done using the DIAGNOSTICS screen "Detector Signal". With zero gas flowing
through the sample cell, adjust the Detector Signal to 0.00
±
0.1 by moving the iris.
See Figure 6-2 for location of iris.
5.3 VOLTAGE CHECKS
Refer to Section 3.8 and verify that the voltages for the detector signal, lamp current,
and three power supplies are correct.
5.4 DIGITAL GAIN ADJUSTMENT
The digitally controlled GAIN amplifier does not normally need adjustment, however, in
the event that the analyzer cannot be spanned, the GAIN must be adjusted as follows:
1. Follow the steps for spanning the analyzer in Sections 3.11 (standard analyzer) or
3.12 (analyzer with the Calibration Gas Control Option) and span the analyzer. If
the Run Mode value is not between 51% and 100% of the span gas value (while
keeping the span potentiometer between 5% and 95%), then the digital GAIN must
be adjusted. (The ideal span potentiometer setting is 50%.) Note the final value of
the PS (potentiometer status) for Step 4.
2. Exit to Run Mode.
3. Follow the steps in Figure 3-6 to obtain the GAIN display in the RANGE parameters
menu.
4. Change the GAIN setting to a value higher or lower than the original value. The
GAIN may be changed to 1, 2, 4, 8, or 16. If the span potentiometer status (PS)
was at the top of its range in Step 2 (95%), then the GAIN should be raised. If the
span potentiometer status (PS) was at the bottom of its range in Step 2 (5%), the
GAIN should be lowered.
Additionally, the HIGH/LOW gain setting may be toggled to achieve an approximate
10:1 ratio.
5. Press SHIFT/ENTER to return to the Run Mode.
6. Repeat Step 1. If the analyzer still cannot be spanned, repeat steps 2 through 5 for
a new GAIN value.
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......