I
NITIAL
S
TARTUP AND
C
ALIBRATION
3-1
January 2000
Rosemount Analytical
748460-A
Prior to shipment this instrument was subjected to extensive factory performance
testing, during which all necessary optical and electrical adjustments were made. The
following instructions are recommended for initial startup and subsequent
standardization of the analyzer.
3.1 LEAK TEST
Perform the Leak Test Procedure in Section 2.8.
3.2 POWER VERIFICATION
1. Verify power switch settings are for available power (115 VAC/230 VAC). Refer to
Section 2.
2. Apply power. On the Power Supply Board, verify that heater LED (D6) is ON.
Refer to Figure 2-1.
F
IGURE
3-1 M
ODEL
890 A
DJUSTMENTS
Power Supply
Board
Option
Boards
Gain Adjust
R3
Display Contrast
Adjust
R8
Signal Board
Isolated Current
Jumper/Test Point
Aperture Tune
Isolated Current
Zero Adjust
R27
Reset
SW1
Micro Board
Isolated Current
Span Adjust
R47
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......