I
NITIAL
S
TARTUP AND
C
ALIBRATION
748460-A
3-3
Rosemount Analytical
January 2000
To activate the manual zero calibration of the analyzer.
To activate the manual span calibration of the analyzer.
To display the configuration and the status of alarms and error messages.
Used in conjunction with left and right or up and down arrows, F1, F2 and
ENTER keys. Pressing the SHIFT key in any display except Run Mode, Zero
Setting, Span Setting and Status causes a
↑
to be displayed at the far right
position. Pressing
→
will then move the cursor 16 characters to the right,
pressing
←
will move the cursor 16 characters to the left, and, if a displayed
parameter is being modified, pressing
↑
will access the highest value
allowed for that parameter and pressing
↓
will access the lowest value
allowed for that parameter.
Software programmable keys for quick access to mode functions. When used
in conjunction with the SHIFT key, two additional functions are available:
SHIFT/F1 and SHIFT/F2. The computer acknowledges the keystrokes by
flashing [** KEY SAVED **] on the display. These four functions can
immediately access a particular display for the following modes: Range,
Diagnostics, Linearization, Alarm, Current Output, Auto Zero/Span or Remote
Range I/O.
To display instrument functions. The standard functions are security, range,
diagnostics, current output and linearization. Additional functions, in
conjunction with option boards, are Auto Zero/Span, Remote Range I/O, and
Alarm.
The up and down arrow keys are used to modify the data in the display.
Depress either the up or down arrow to change the values displayed. When
used in one of the editing modes, SHIFT
↑
causes the highest value
allowed in a function to be displayed. SHIFT
↓
causes the lowest value to be
displayed. Depress the arrow key once to change one digit; depress and hold
either key to scroll (continuous value change), thereby reducing the time
required to make large value changes.
To move cursor one position at a time or, when used in conjunction with
the SHIFT key moves the cursor 16 characters, one full display, at a time.
ZERO
SPAN
F1
F2
STATUS
SHIFT
MODE
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......