I
NITIAL
S
TARTUP AND
C
ALIBRATION
748460-A
3-21
Rosemount Analytical
January 2000
3.14.1 STATUS D
ISPLAY
Refer to Figure 3-13. The STATUS display shows the alarms, error messages and
security lockout status. (See explanations of error messages in Section 5.1.) The
STATUS display can be used to check the following alarm setpoints without entering
one of the MODE functions: HIGH/LOW, AUTO/MANUAL and ON/OFF.
The order of priority for error messages, security status and alarms is as follows:
[SECURITY ENABLED/DISABLED]
[ERL ]
[EL-LIN.COEFF ERR]
[E0-ZERO POT LMTS]
[E1-R1 SP POT LMTS]
[E2-R2 SP POT LMTS]
[E3-R3 SP POT LMTS]
[E4-ADC SATURATED]
[E5-ZERO DRIFT]
[E6-SPAN DRIFT]
[E7-GAIN TOO HIGH]
[RMT: R/L]
[ALARM 1 AUTO/{MAN ON/OFF}]
[ALARM 2 AUTO/{MAN ON/OFF}]
[CAL-CTL PRESENT]
[AUTOCAL: ON/OFF]
[CURRENT 0/4 SP ON/OFF]
F
IGURE
3-13. S
TATUS
D
ISPLAY
Alarm:
MN (Manual) or
AT (Automatic)
Alarm 1 (15 sec.)
{Followed by]
Alarm 2 (15 sec.)
Alarm ON
or OFF
ALARM 1 MN OFF
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......