![Rosemount 890 Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/rosemount/890/890_instruction-manual_1487417043.webp)
I
NITIAL
S
TARTUP AND
C
ALIBRATION
748460-A
3-13
Rosemount Analytical
January 2000
3.10 ZERO CALIBRATION FOR THE ANALYZER WITH THE CAL
GAS CONTROL OPTION
The Calibration Gas Control Option allows one-man calibration. This option consists
of two form C contacts, rated 3A-125/250 VAC or 5A-30 VDC (resistive). These
contacts are connected to solenoid valves (customer supplied) which will turn zero and
span calibration gases ON/OFF when activated. Simply press ZERO to open the
appropriate valve, thus allowing the zero gas to flow through the instrument. Pressing
ENTER will initiate the calibration process. At the conclusion of calibration, the valve
is closed and the instrument returns to RUN mode.
Note
For instruments with Cal Gas Option or Auto Zero/Span Option, press ENTER to
perform the selected calibration. Press SHIFT-ENTER to abort the function, turn
off the relay for this valve, and maintain previous calibration settings.
Note
When entering this function for viewing purposes only (by pressing the ZERO or
SPAN key), press SHIFT-ENTER to exit.
To calibrate the analyzer with the Cal Gas Control Option:
1. Allow system to warm up a minimum of one hour.
2. Connect the solenoid valve for the zero gas to the two form C contacts. Connect
the zero gas to the sample cell inlet located on the back of the analyzer. The gas
should flow at a rate of 500 cc/min., as read on a flowmeter, for at least two
minutes.
3. To calibrate the zero setting for the analyzer, press ZERO and then ENTER.
3.11 SPAN CALIBRATION
The Model 890 Analyzer sets span for the selected range in a simple two-keystroke
procedure.
To span the analyzer, the operator simply presses SPAN and ENTER. When SPAN is
pressed, display notes [SPAN - WAIT] and then [X.XX NN % MMM
→
], where X.XX
is the run mode value, NN % is the percentage of span potentiometer in use and MMM
is the span gas value. When ENTER is pressed, display notes [CALCULATING SPAN]
and the analyzer re-enters Run mode on completion of function.
Note: Press SHIFT-ENTER instead of ENTER to maintain previous calibration
settings and exit Span without calibrating.
Note: For instruments with the Calibration Gas Control or Auto Zero/Span,
press ENTER to perform the new calibration. Both ENTER and SHIFT/ENTER
will turn off the relay for the solenoid valve.
Содержание 890
Страница 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Страница 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Страница 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Страница 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......