158
September 2017
Reference Manual
00809-0100-4803, Rev GA
Special Conditions for Safe Use (X):
1. Cable entries must be used which maintain the ingress
protection of the enclosure to at least IP66.
2. Unused cable entries must be filled with suitable blanking
plugs which maintain the ingress protection of the enclosure
to at least IP66.
3. Cable entries and blanking plugs must be suitable for the
ambient temperature range of the apparatus and capable of
withstanding a 7J impact test.
4. The Rosemount 3051S - SuperModule must be securely
screwed in place to maintain the ingress protection of the
enclosure.
I7
IECEx Intrinsic Safety
Certificate: IECEx BAS 08.0025X
Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ia IIC T4 Ga, T4(–60 °C
≤
T
a
≤
+70 °C)
Special Conditions for Safe Use (X):
1. If the equipment is fitted with the optional 90 V transient
suppressor, it is incapable of withstanding the 500 V isolation
from earth test and this must be taken into account during
installation.
2. The enclosure may be made of aluminum alloy and given a
protective polyurethane paint finish; however, care should
be taken to protect it from impact or abrasion if located in a
Zone 0 environment.
IG
IECEx FISCO
Certificate: IECEx BAS 08.0025X
Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ia IIC T4 Ga, T4(–60 °C
≤
T
a
≤
+70 °C)
N7
IECEx Type n
Certificate: IECEx BAS 08.0026X
Standards: IEC 60079-0: 2011, IEC 60079-15: 2010
Markings: Ex nA IIC T5 Gc, (–40 °C
≤
T
a
≤
70 °C)
Special Condition for Safe Use (X):
1. If fitted with a 90 V transient suppressor, the equipment is
not capable of withstanding the 500 V electrical strength
test as defined in Clause 6.5.1 of IEC 60079-15:2010. This
must be taken into account during installation.
B.8 Brazil
E2
INMETRO Flameproof
Certificate: UL-BR 15.0393X
Standards: ABNT NBR IEC 60079-0:2008 + Corrigendum
1:2011,
ABNT NBR IEC 60079-1:2009 + Corrigendum
1:2011,
ABNT NBR IEC 60079-26:2008 + Corrigendum 1:
2008
Markings: Ex d IIC T* Ga/Gb, T6(–60 °C
≤
T
a
≤
+70 °C),
T5/T4(-60 °C
≤
T
a
≤
+80 °C), IP66
Special Conditions for Safe Use (X):
1. The device contains a thin wall diaphragm. Installation,
maintenance and use shall take into account the
environmental conditions to which the diaphragm will be
subjected. The manufacturer's instructions for installation
and maintenance shall be followed in detail to assure safety
during its expected lifetime.
2. For information on the dimensions of the flameproof joints,
the manufacturer shall be contacted.
I2
INMETRO Intrinsic Safety
Certificate: UL-BR 15.0357X
Standards: ABNT NBR IEC 60079-0:2008 + Addendum
1:2011, ABNT NBR IEC 60079-11:2009
Markings: Ex ia IIC T4 Ga (–60 °C
≤
T
a
≤
+70 °C)
Parameter
HART
F
OUNDATION
Fieldbus
SuperModule
only
Voltage U
i
30 V
30 V
7.14 V
Current I
i
300 mA
300 mA
300 mA
Power P
i
1 W
1.3 W
887 mW
Capacitance C
i
14.8 nF
0
0.11
μ
F
Inductance L
i
0
0
0
Parameter
RTD (for 3051SFx)
(HART)
RTD (for 3051SFx)
(Fieldbus)
Voltage U
i
30 V
30 V
Current I
i
2.31 mA
18.24 mA
Power P
i
17.32 mW
137 mA
Capacitance C
i
0
0.8 nF
Inductance L
i
0
1.33 mH
Parameter
FISCO
Voltage U
i
17.5 V
Current I
i
380 mA
Power P
i
5.32 W
Capacitance C
i
0
Inductance L
i
0
Содержание 3051S Series
Страница 1: ...Reference Manual 00809 0100 4803 Rev GA September 2017 Rosemount 3051S MultiVariable Transmitter ...
Страница 2: ......
Страница 114: ...106 Reference Manual 00809 0100 4803 Rev GA Operation and Maintenance September 2017 Operation and Maintenance ...
Страница 126: ...118 Reference Manual 00809 0100 4803 Rev GA Troubleshooting September 2017 Troubleshooting ...
Страница 171: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 163 Product Certifications ...
Страница 173: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 165 Product Certifications ...
Страница 176: ...168 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 180: ...172 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA CO S CO O ...
Страница 181: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 173 Product Certifications ...
Страница 182: ...174 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 184: ...176 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 185: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 177 Product Certifications ...
Страница 187: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 179 Product Certifications ...
Страница 188: ...180 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 189: ......