Specifications and Reference Data
September 2017
Reference Manual
00809-0100-4803, Rev GA
Specifications and Reference Data
126
Rosemount
models
Ultra
Classic
Ultra for Flow
3051SMV_ _3: Differential pressure and temperature
3051SMV_ _4: Differential pressure
Ranges 2–4
±0.025% of span;
For spans less than 10:1,
±0.055% of span;
For spans less than 10:1,
±0.04% of reading up to 8:1
DP turndown from URL;
±(0.04 + 0.0023[URL/RDG
])%
reading to 200:1 DP turndown
from URL
Range 5
±0.05% of span;
For spans less than 10:1,
±0.065% of span;
For spans less than 10:1,
N/A
Range 1
±0.09% of span;
For spans less than 15:1,
±0.10% of span;
For spans less than 15:1,
N/A
Range 0
±0.09% of span;
For spans less than 2:1,
±0.045% of URL
±0.10% of span;
For spans less than 2:1,
±0.05% of URL
N/A
Process Temp.
RTD Interface
±0.67 °F (0.37 °C)
±0.67 °F (0.37 °C)
±0.67 °F (0.37 °C)
1.
Stated reference accuracy equations include terminal based linearity, hysteresis, and repeatability, but does not include analog only reference accuracy of ±0.005% of
span.
2.
RDG refers to transmitter DP reading.
3.
Ultra for Flow is only available for Rosemount
™
3051SMV DP Ranges 2–3. For calibrated spans from 1:1 to 2:1 of URL, add ±0.005% of span analog output error.
4.
Specifications for process temperature are for the transmitter portion only. The transmitter is compatible with any Pt 100 (100 ohm platinum) RTD. Examples of
compatible RTDs include Rosemount Series 68 and 78 RTD Temperature Sensors.
0.005
0.0035
+
URL
span
-------------
% of span
±
0.015
0.005
+
URL
span
-------------
% of span
±
0.005
0.0045
+
URL
span
-------------
% of span
±
0.015
0.005
+
URL
span
-------------
% of span
±
0.015
0.005
+
URL
span
-------------
% of span
±
0.025
0.005
+
URL
span
-------------
% of span
±
Total performance
(1)
Models
Ultra
Classic and Classic MV
Ultra for Flow
(2)
Rosemount
3051SMV
DP Ranges 2–3
±0.1% of span;
for ±50 °F (28 °C) temperature
changes; 0-100% relative
humidity, up to 740 psi (51 bar)
line pressure (DP only),
from 1:1 to 5:1 rangedown
±0.15% of span;
for ±50 °F (28 °C) temperature
changes; 0-100% relative
humidity, up to 740 psi (51 bar)
line pressure (DP only),
from 1:1 to 5:1 rangedown
±0.1% of reading;
for ±50 °F (28 °C) temperature
changes; 0-100% relative
humidity, up to 740 psi (51 bar)
line pressure,
over 8:1 DP turndown from URL
1.
Total performance is based on combined errors of reference accuracy, ambient temperature effect, and line pressure effect. Specifications apply only to differential
pressure measurement.
2.
Ultra for Flow is only available for Rosemount 3051SMV DP Ranges 2–3.
Multivariable flow performance
(1)
Mass, energy, actual volumetric, and totalized flow reference accuracy
(2)
Models
Ultra for Flow
Classic MV
Rosemount
3051SMV
DP Ranges 2–3
±0.65% of flow rate over a 14:1 flow range
(200:1 DP range)
±0.70% of flow rate over 8:1 flow range
(64:1 DP range)
DP Range 1
N/A
±0.90% of flow rate over 8:1 flow range
(64:1 DP range)
1.
Applies to the 3051SMV_M Multivariable
™
type only. Flow performance specifications assume device is configured for full compensation of static pressure, process
temperature, density, viscosity, gas expansion, discharge coefficient, and thermal correction variances over a specified operating range.
2.
Uncalibrated differential producer (0.2 < beta < 0.6 Orifice) installed per ASME MFC 3M or ISO 5167-1. Uncertainties for discharge coefficient, producer bore, tube
diameter, and gas expansion factor as defined in ASME MFC 3M or ISO 5167-1. Reference accuracy does not include RTD sensor accuracy.
Содержание 3051S Series
Страница 1: ...Reference Manual 00809 0100 4803 Rev GA September 2017 Rosemount 3051S MultiVariable Transmitter ...
Страница 2: ......
Страница 114: ...106 Reference Manual 00809 0100 4803 Rev GA Operation and Maintenance September 2017 Operation and Maintenance ...
Страница 126: ...118 Reference Manual 00809 0100 4803 Rev GA Troubleshooting September 2017 Troubleshooting ...
Страница 171: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 163 Product Certifications ...
Страница 173: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 165 Product Certifications ...
Страница 176: ...168 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 180: ...172 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA CO S CO O ...
Страница 181: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 173 Product Certifications ...
Страница 182: ...174 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 184: ...176 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 185: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 177 Product Certifications ...
Страница 187: ...Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA 179 Product Certifications ...
Страница 188: ...180 Product Certifications Product Certifications September 2017 Reference Manual 00809 0100 4803 Rev GA ...
Страница 189: ......