Manual
Double Sheet Control System R1000 series L20
with integrated fieldbus interface
Start-up
B0063191 / 1.3
80
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
For L20-XX-S:
Adjustment parameter 4:
Calibration deviation
Fig. 51: Selection of the calibration deviation
The measured deviation of the calibration sheet thickness (in mm / inch) is displayed by pressing
the ENTER key. This requires that the calibration sheet has to be inserted.
Fig. 52: Display of the calibration deviation
Pressing the ENTER key will close the display of the calibration deviation.
For L20-3-XX-S:
Adjustment parameter 4:
System test / Alignment
According to the type of sensor used, the device performs the following check:
A)
When connecting the sensor type , (in preparation) and
(in preparation) the
alignment
and
function control
of
all
activated sensors in the system settings
is performed.
Fig. 53: Display of system test / alignment
Possible selections:
yes
no
adjust parameter: 4
show calibration
deviation distance
adjust parameter: 4
AB1:-0.005 mm
adjust parameter: 4
system test/
alignment
no