Manual
Double Sheet Control System R1000 series L20
with integrated fieldbus interface
Start-up
B0063191 / 1.3
78
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
L20-XX-S
L20-3-XX-S
Fig. 46: Feed the calibration sheet
L20-XX-S:
At this point place the calibration sheet in the measuring station. The calibration is
performed punctual. During the calibration a sheet
must
be between the sensors.
L20-3-XX-S:
Now the calibration sheet is placed in the measuring station or passed through the
measuring station. Confirm by pressing the ENTER key.
Attention
For L20-3-XX-S dynamic calibration is advantageous because
localized thickness variations or fluctuations in the calibration piece
will have less impact on the calibration.
Note
Occurring fault messages read section „9. Fault messages“.
Now the rated sensor range is shown. The margin areas must be avoided.
Fig. 47: Showing of the rated sensor range
If several measuring points have been calibrated, the ABs will be listed consecutively in the last
row. The measuring channel calibration is completed by pressing the ENTER key.
Adjustment parameter 2:
Controlling the sensor gaps with the calibration sheet.
Fig. 48: Select the control with the calibration sheet
To perform a control press the ENTER key.
Selection options:
•
no:
Do not perform control.
•
yes:
Perform
control
(At
L20-XX-S)
•
measuring channel:
1 / 2 / 3 / 1+2 / 1+3 / 2+3 / 1+2+3
perform controls at selected measuring channels
(At L20-3-XX-S and depending on connected sensors)
adjust parameter: 1
feed calibration
sheet at channel
push ENTER
adjust parameter: 1
feed calibration
active
push ENTER to finish
adjust parameter: 1
rated sensor range
37.5 ... 42.5mm
AB1 active: 41mm
adjust parameter: 2
control with
calibration sheet
no