R&S FSQ
Preparing for Use
LXI Configuration
2.42
Quick Start Guide 1313.9652.62 - 02
2.6.3 LAN Configuration website
The LAN Configuration web page displays all mandatory LAN parameters and allows their
modification.
The TCP/IP Mode configuration field controls how the IP address for the instrument gets
assigned (see also <Link to IP Address Settings>). For the manual configuration mode, the
static IP address, subnet mask, and default gateway are used to configure the LAN. The
automatic configuration mode uses DHCP server or Dynamic Link Local Addressing
(Automatic IP) to obtain the instrument IP address.
The links across the bottom of the LAN Configuration page open other pages:
Ping Client provides the ping utility to verify the connection between the instrument and other
devices.
Advanced LAN Configuration provides LAN settings that are not declared mandatory by the
LXI standard.
The Advanced LAN Configuration parameters are used as follows:
• The Negotiation configuration field provides different Ethernet speed and duplex mode
settings. In general, the Auto Detect mode is sufficient.
• ICMP Ping must be enabled to use the ping utility.
• VXI-11 is the protocol that is used for discovery of the instrument in the LAN. According to
the standard, LXI devices must use VXI-11 to provide a discovery mechanism; other
additional discovery mechanisms are permitted.
2.6.4 Ping
Ping is a utility that verifies the connection between the LXI-compliant instrument and another
device. The ping command uses the ICMP echo request and echo reply packets to determine
whether the LAN connection is functional. Ping is useful for diagnosing IP network or router
failures.
The ping utility is not password-protected. To initiate a ping between the LXI-compliant
instrument and a second connected device,
1. Enable ICMP Ping on the Advanced LAN Configuration page (enabled after an LCI).
2. Enter the IP address of the second device without the ping command and without any
further parameters into the Destination Address field (e.g. 10.113.10.203).
3. Click Submit.
Changing the LAN configuration is password-protected. The password
reads LxiWebIfc (notice upper and lower case characters).
Содержание FSQ series
Страница 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Страница 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Страница 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Страница 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Страница 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Страница 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Страница 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...