Quick Start Guide 1313.9652.62 - 02
5.13
R&S FSQ
Measuring Harmonics with Frequency Sweeps
5. Set the measured signal frequency and the measured level as reference values.
➢
Press the
REFERENCE FIXED
softkey.
The position of the marker becomes the reference point. The reference point level is
indicated by a horizontal line, the reference point frequency with a vertical line. At the
same time, the delta 2 marker is switched on at the marker position.
Fig. 5-4
Fundamental wave and the frequency and level reference point
6. Make the step size for the center frequency equal to the signal frequency.
➢
Press the
FREQ
key.
➢
Press the
CF STEPSIZE
softkey and press the
=MARKER
softkey in the submenu.
The step size for the center frequency is now equal to the marker frequency.
7. Set the center frequency to the 2
nd
harmonic of the signal
➢
Press the
FREQ
key.
➢
Press the up arrow key once.
The center frequency is set to the 2
nd
harmonic.