Quick Start Guide 1313.9652.62 - 02
5.31
R&S FSQ
Measurements in the Time Domain
3. Set the resolution bandwidth to 300 kHz.
➢
Press the
BW
key.
➢
Press the
RES BW MANUAL
softkey and enter
300 kHz
.
4. Set the display range to 20 dB and shift the filter trace to the center of the screen.
➢
Press the
AMPT
key.
➢
Press the
RANGE LOG MANUAL
softkey and enter
20 dB
.
➢
Press the
NEXT
key.
➢
Set the
GRID
softkey to
REL
.
➢
Press the
PREV
key.
➢
Using the rotary knob, set the reference level such that the filter edge at the center
frequency intersects the –10 dB level line.
The filter edge of the 300 kHz filter will appear on screen. This corresponds to the
demodulator characteristic for FM signals with a steepness of approx. 5 dB/100 kHz.
Fig. 5-17 Display of the filter edge of the 300 kHz filter as an FM discriminator
characteristic
5. Set the FM deviation to 100 kHz and the AF to 1 kHz on the signal generator.
6. Set the frequency deviation to 0 Hz on the R&S FSQ.
➢
Press the
SPAN
key.
➢
Press the
ZERO SPAN
softkey.
The demodulated FM signal will appear on screen. The signal is continuous across the
screen.
Содержание FSQ series
Страница 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Страница 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Страница 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Страница 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Страница 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Страница 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Страница 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...