Quick Start Guide 1313.9652.62 - 02
5.21
R&S FSQ
Measurements in the Time Domain
5.5 Measurements in the Time Domain
In the case of radio transmission systems that use the TDMA method (e.g. GSM or IS136),
transmission quality is determined not only by spectral characteristics but also by
characteristics in the time domain. However, a timeslot is assigned to each user since several
users share the same frequency. Smooth operation is ensured only if all users adhere exactly
to their assigned timeslots.
Both the power during the send phase as well as the timing and duration of the TDMA burst
and rise and fall times of the burst are important.
5.5.1 Measuring the Power Characteristic
For measuring power in the time domain, the R&S FSQ offers easy-to-use functions that
measure the power over a predefined time.
5.5.1.1 Example – Measuring the Power of a GSM Burst During the Activation Phase.
Test setup:
Settings on the signal generator (e.g. R&S SMIQ):
Measurement with the R&S FSQ:
1. Reset the instrument.
➢
Press the
PRESET
key.
The R&S FSQ is now in the default state.
2. Set the center frequency to 100 MHz, the span to 0 Hz and the resolution bandwidth
to 1 MHz.
➢
Press the
FREQ
key and enter
100 MHz
.
➢
Press the
SPAN
key and enter
0 Hz
, or press the
ZEROSPAN
softkey.
3. Set the reference level of the R&S FSQ to 10 dBm (= level of the signal generator
+10 dB).
➢
Press the
AMPT
key and enter
10 dBm
.
Frequency:
100 MHz
Level:
0 dBm
Modulation:
GSM, one timeslot activated
Signal
generator
R&S FSQ
Содержание FSQ series
Страница 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Страница 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Страница 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Страница 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Страница 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Страница 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Страница 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...