Remote Commands
R&S
®
EVSG1000
270
User Manual 1178.6227.02 ─ 06
Manual operation:
See
10.1.11
Network Settings and Remote Operation
The following commands are required to configure network settings and remote opera-
tion.
GETHOSTNAME
This command returns the host name of the instrument.
Return values:
<hostname>
Manual operation:
See
LO
This command sets the R&S
EVSG1000 back to local control.
Return values:
<State>
READY.
The command was executed successfully.
Usage:
Setting only
MLOC
The
MLOC
command is identical to the
LO
command.
Return values:
<State>
READY.
The command was executed successfully.
RLC
<State>
REMOTELOCK
<State>
This command deactivates/activates local operation by keyboard.
Parameters:
<State>
ON
If set, no local operation by keyboard is possible.
The "ESC" button has no effect.