Graphic Measurement Modes
R&S
®
EVSG1000
165
User Manual 1178.6227.02 ─ 06
Defines the reference power in volt for LF input, available for AF signals with AC cou-
pling only. This value is used to determine the modulation depth and corresponds to
the DC power of the AF signal.
For AC coupling, the DC power component of the AF signal is not forwarded. Thus,
you must define the reference value manually.
Remote command:
LF In Factor
Access
: [AMPT]
Access VNC
: [w]
Applies a factor to the measured power levels, e.g. to compensate certain effects in the
input system.
Remote command:
5.4.5
Configuring the Measurement Time
Access
: [MTime] > "Time/Div"
Access VNC
: [d] > "F1"
Time/Div
Defines the time range displayed in each division of the x-axis. The full x-axis range,
that is: the total measurement time, is eight times this value.
Remote command:
5.4.6
Configuring the Trigger
Access
: [Trigger]
Access VNC
: [k]
The trigger determines how often and when a new measurement is performed.
Performing a single sweep
To perform a single sweep, press [Single] (VNC: [L]).
AF Time Domain Mode (Option R&S