GBAS/SCAT-I Mode (Options R&S
EVSG-K4/-K5)
R&S
®
EVSG1000
188
User Manual 1178.6227.02 ─ 06
The individual measurement results are described in
Chapter 6.3.2.2, "Frame Results
Recording GBAS/SCAT-I measurement results
When recording data in GBAS/SCAT-I mode, one record is stored for each slot, with a
duration of 62.5
ms. The record contains measurement values and message data.
Recording is continued even if the display update is temporarily stopped (by the
"Pause" function, see
on page 190). Start and stop recording from any
measurement view using the [Record] key. Record only the selected and currently dis-
played slot or frame using the [Single] key.
You can define which data is stored; see
Chapter 8.2.3, "Recording Settings"
For details on data logging, see
Chapter 8.2, "Recording Measurement Data"
GBAS / SCAT-I Measurement Views and Results