Getting Started
R&S
®
EVSG1000
21
User Manual 1178.6227.02 ─ 06
1.4.4
Setting Up the R&S
EVSG1000
The R&S
EVSG1000 is designed for use either on a bench top or in a rack, or as a
portable instrument (with optional battery operation) in a transport bag in the field.
See also:
●
●
1.4.4.1
Placing the R&S
EVSG1000 on a Bench Top
To place the R&S
EVSG1000 on a bench top
1. Place the R&S
EVSG1000 on a stable, flat and level surface. Ensure that the sur-
face can support the weight of the instrument. For information on the weight, see
the data sheet.
2.
CAUTION!
The instrument can fall over and cause injury. The top surface is not
designed for stacking. Never stack another instrument on top of the
R&S
EVSG1000.
As an alternative, you can mount several instruments in a rack.
3.
NOTICE!
Overheating can damage the product.
Prevent overheating as follows:
● Keep a minimum distance of 10
cm between the fan openings of the
R&S
EVSG1000 and any object in the vicinity.
● Do not place the R&S
EVSG1000 next to heat-generating equipment such as
radiators or other products.
1.4.4.2
Mounting the R&S
EVSG1000 in a Rack
A 19"-adapter is available as an accessory (R&S
EVSG-Z7) to mount the
R&S
EVSG1000 in a 19'' rack (e.g. in a measurement vehicle).
To prepare the rack
1. Observe the requirements and instructions in
Preparing for Use