Numeric Measurement Modes
R&S
®
EVSG1000
84
User Manual 1178.6227.02 ─ 06
Remote command:
Low IF Offset
Access
: [Config]
Access VNC
: [s]
Defines the frequency of the low IF input signal, that is: the offset from 0
Hz.
Only available for low IF mode (see
Remote command:
Range
Access
: [AMPT]
Access VNC
: [w]
Defines the full scale AF signal level. Define the correct level to avoid clipping and
overload.
Switches the range between 1
V and 5
V.
Remote command:
Coupling
Access
: [AMPT]
Access VNC
: [w]
Specifies the handling of the DC component of the AF signal.
For low IF signals, AC coupling is always used.
"AC"
The DC component of the AF signal is not forwarded.
"DC"
The DC component of the AF signal is also forwarded and analyzed.
Remote command:
DC Reference
Access
: [AMPT]
Access VNC
: [w]
Defines the reference power in volt for LF input, available for AF signals with AC cou-
pling only. This value is used to determine the modulation depth and corresponds to
the DC power of the AF signal.
For AC coupling, the DC power component of the AF signal is not forwarded. Thus,
you must define the reference value manually.
Remote command:
ILS Localizer and ILS Glidepath (GP) Modes