CTS60
Module Test
1094.0006.60
11.5
E-4
Exp Power
Setting of the maximum expected power between -10 dBm and 30 dBm in steps of 1 dB for fully utilising
the dynamic range of the test receiver of the CTS60.
Note:
If a value below the power of the measured signal is selected, the measured NTP value is not correct.
RF Chan
Input of the channel number of the signal to be received. Channel numbers between -3 and 12 at half
the DECT channel spacing (864 kHz) are possible.
Config
Pressing the “Config” softkey allows some of the static configuration information displayed on this menu
to be modified. The modification of this configuration information is described in section 11.1.
The following static configuration information is displayed to which the following explanations apply.
Input. Att.
The RF input connector correction value for external path attenuation currently in use, is displayed in
this field.
Trigger Mode
•
Free Run. No trigger condition is currently present.
•
Rising Slope. The rising power is the current trigger event.
Trigger Delay
This is the measurement delay after the trigger event from 0 to 10 ms in steps of 1 µs that is currently in
use. If the trigger mode is set to “Free Run”, then the value set for “Trigger Delay” is irrelevant. “Trigger
Delay” can also be used to display the rising edge of a periodical signal triggered with the condition
“Rising Slope”.
Meas. Window
The currently selected measurement window of discrete values between 25 µs and 1000 µs
Mod.Ref
The current reference frequency against which the modulation is measured. The values are:
•
Measured. The mean frequency of the measured signal. This mode is also used in manual test.
•
Nominal. The exact centre frequency of the carrier, as selected by the user.
•
Half Peak. The mid-point between the measured minimum and maximum measured peaks.