CTS60
Autotest Printed Reports
1094.0006.60
7.9
E-3
7.3.5
Timing (Portable Part)
____________________________________________________________________
| (03) Timing (Bursts = 32): |
| Value Limits Result |
| Min. Max. |
| Max. Pos. Jitter (us) 0.03 1.0 OK |
| Max. Neg. Jitter (us) -0.84 -1.0 OK |
| Max. Packet Delay (us) 0.53 2.0 OK |
| Min. Packet Delay (us) -0.31 -2.0 OK |
|____________________________________________________________________|
This section of the report displays the results of the portable part timing measurement. The first line
details the test step number in brackets, and the title "Timing" to show that this is a timing measurement.
After this, in brackets, is the number of bursts configured for this measurement.
Below this is a table of the configured values and measured results for the measurements. The first
column details what was measured and its unit in brackets. Under "Value" is listed all of the measured
values for this measurement. The column "Limits" displays the configured tolerance limits for each of
the measurements. The final column "Result" gives a result as to whether the measurement has passed
the tolerance check.
Max. Pos. Jitter
This line details the maximum positive jitter value measured for this test step in
µ
s. To be in tolerance
this value must be less than the limit specified under the "Limits" column.
Max. Neg. Jitter
This line details the maximum negative jitter value measured for this test step in
µ
s. To be in tolerance
this value must be greater than the limit specified under the "Limits" column.
Max. Packet Delay
This line details the maximum packet delay value measured for this test step in
µ
s. To be in tolerance
this value must be less than the limit specified under the "Limits" column.
Min. Packet Delay
This line details the minimum packet delay value measured for this test step in
µ
s. To be in tolerance
this value must be greater than the limit specified under the "Limits" column.