![RKI Instruments GD-70D Series Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/rki-instruments/gd-70d-series/gd-70d-series_operating-manual_1466822088.webp)
GD-70D
- 88 -
Detection Gas
Range
Calibration Instructions
COF2 (Cobalt difluorine)
0 - 6.00 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
DEA (Diethylamine)
0 - 15.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
DMA (Dimethylamine)
0 - 15.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
F2 (Fluorine)
0 - 1.00 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
0 - 3.00 ppm
GeH4 (Germane)
0 - 0.80 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
H2 (Hydrogen)
0 - 100 %LEL
Calibrating with a Demand Flow Regulator on page 90
0 - 2.00 %volume
Calibrating with a Humidifier (MOS Sensors) on page 103
0 - 500 ppm
0 - 1,000 ppm
0 - 2,000 ppm
0 - 4,000 ppm
H2Cl2Si (Dichlorosilane)
0 - 15.0 ppm
Calibrating with a Gas Bag (Calibrating with HCl) on page 99
H2O2 (Hydrogen peroxide)
0 - 3.00 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
H2S (Hydrogen sulfide)
0 - 1.00 ppm
Calibrating with a Demand Flow Regulator on page 90
0 - 30.0 ppm
H2Se (Hydrogen selenide)
0 - 0.20 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
H9NSi3 (Trisilylamine)
0 - 15.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
HBR (Hydrogen bromide)
0 - 6.00 ppm
Calibrating with a Gas Bag (Calibrating with HCl) on page 99
0 - 9.00 ppm
HCl (Hydrogen chloride)
0 - 6.00 ppm
Calibrating with a Gas Bag (Calibrating with HCl) on page 99
0 - 15.0 ppm
HCN (Hydrogen cyanide)
0.9 - 15.0 ppm
Calibrating with a Demand Flow Regulator on page 90
HCOOH (Formic acid)
0 - 15.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
HF (Hydrogen fluoride)
0.4 - 3.00 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
0.4 - 9.00 ppm
HNO3 (Nitric acid)
0 - 20.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93