- 53 -
GD-70D
6-5
About LONWORKS (LN specification)
<
Binding method
>
≪
User mode
≫
PW
A1
A2
F
LCD
Detection mode
Press the MODE key
for 3 seconds.
●
○
○
○
0.0ppm
SIH4
↓
User mode
↓
1-1.ZERO
Perform the zero
adjustment.
●
○
○
○
1- 1
ZERO
MAINTENANCE
→
SET
Zero Adjustment
⇒
P51
To reissue the neuron ID, please press the MODE key in user mode for 3 seconds, return to the detection
mode, and enter the user mode again.
WARNING
•
When adjustment is completed, press the MODE key and be sure to return to detection
mode.
(
When left in user mode, it automatically returns to detection mode after 10 hours.
)